Phys Med Biol. 2002 Nov 21;47(22):3959-72.

CdZnTe detector in mammographic x-ray spectroscopy.

Author information

1
Graduate School of Medicine, Course of Health Sciences, Osaka University, Suita, Japan.

Abstract

A CdZnTe (CZT) detector was utilized in mammographic x-ray spectroscopy under clinical conditions. First, the detector response was investigated using y-rays from 241Am. The escape of secondary (Compton scattered and K fluorescent) x-rays and tailing due to carrier trapping were minor in the mammographic energy range. In addition, the transmission of primary x-rays was minimal from the results calculated using the mass attenuation coefficients of CZT. Therefore, spectral distortion in this energy range was expected to be negligible. Secondly, x-ray spectroscopy was carried out with the CZT detector. The measured spectra were in good agreement with the spectra obtained with the Compton-scatter method with a high-purity germanium detector. Moreover, the half-value layers (HVLs) calculated from the CZT spectra were consistent with the HVLs measured with an ionization chamber. The results indicate that a CZT detector can be utilized in mammographic x-ray spectroscopy without any corrections.