Format

Send to

Choose Destination
J Biomed Opt. 2002 Oct;7(4):643-9.

Analysis on performance and optimization of frequency-domain near-infrared instruments.

Author information

1
University of Pennsylvania, Department of Electrical Engineering, Department of Biochemistry and Biophysics, Philadelphia, Pennsylvania 19104, USA. tutao@mail.upenn.edu

Abstract

Frequency-domain near-infrared techniques have been widely used to detect the optical properties of biological tissues noninvasively. In this paper we propose an analytical model to evaluate the performance of frequency-domain instruments. Based on the diffusion equation and the transfer properties of optoelectronic components, we treat all parts, including the medium, as two-port networks and apply systematic methods to answer questions concerning frequency-domain instruments. Experiments show that this method can reasonably reflect the properties of the instrument within an accuracy of 7%. This kind of method can be used to design suitable instruments for various applications. We also analyze the selection of the instrument parameters to achieve optimal performance at an efficient cost using this analytical model.

PMID:
12421133
DOI:
10.1117/1.1501562
[Indexed for MEDLINE]