Influence of readout parameters on TL response, re-usability and residual signal in LiF:Mg,Cu,P

Radiat Prot Dosimetry. 2002;100(1-4):353-6. doi: 10.1093/oxfordjournals.rpd.a005887.

Abstract

It has recently been recommended that heating rates do not exceed 10 K.s(-1) and that the maximum temperature of readout should not exceed 265 degrees C for LiF:Mg,Cu,P. In some cases, a decrease of sensitivity in this material in the first of several re-use cycles had been reported. Influence of heating rates up to 30 K.s(-1), duration time up to 40s and maximum readout temperatures up to 270 degrees C on TL response, re-usability and residual signal was investigated. It was found that the maximum readout temperatures above 240 degrees C may lead to the thermoluminscent response decrease in the first several re-use cycles. The readout parameters can be optimised to minimise the residual signal (less than 0.4%) and to retain a constant sensitivity at the same time at high heating rates up to 30 K.s(-1) in a short time (less than 1 min per TL chip) without the necessity of heating above 240 degrees C. A concept of 'efficient residual signal' was put forward to quantify more accurately the real residual signal which affects the precision of the next measurement.

MeSH terms

  • Copper / chemistry
  • Fluorides / chemistry
  • Fluorides / radiation effects*
  • Hot Temperature
  • Lithium Compounds / chemistry
  • Lithium Compounds / radiation effects*
  • Luminescent Measurements
  • Magnesium / chemistry
  • Phosphorus / chemistry
  • Radiochemistry
  • Sensitivity and Specificity
  • Thermoluminescent Dosimetry / methods*
  • Thermoluminescent Dosimetry / statistics & numerical data

Substances

  • Lithium Compounds
  • lithium fluoride
  • Phosphorus
  • Copper
  • Magnesium
  • Fluorides