In vivo real-time intravascular MRI

J Cardiovasc Magn Reson. 2002;4(2):223-32. doi: 10.1081/jcmr-120003948.

Abstract

Purpose: The Magnetic resonance imaging (MRI) is an emerging technology for catheter-based imaging and interventions. Real-time MRI is a promising methodfor overcoming catheter and physiologic motion for intravascular imaging.

Methods: All imaging was performed on a 1.5 T Signa MRI scanner with high-speed gradients. Multiple catheter coils were designed and constructed, including low-profile, stub-matched coils. Coil sensitivity patterns and SNR measurements were compared. Real-time imaging was performed with an interleaved spiral sequence using a dedicated workstation, providing real-time data acquisition, image reconstruction and interactive control and display. Real-time "black-blood" imaging was achieved through incorporation of off-slice saturation pulses. The imaging sequence was tested in a continuous flow phantom and then in vivo in the rabbit aorta using a 2 mm catheter coil.

Results: The real-time intravascular imaging sequence achieved 120-440 micron resolution at up to 16 frames per second. Low-profile stub-tuned catheter coils achieved similar SNR to larger traditional coil designs. In the phantom experiments, addition of real-time black-blood saturation pulses effectively suppressed the flow signal and allowed visualization of the phantom wall. In vivo experiments clearly showed real-time intravascular imaging of the rabbit aortic wall with minimal motion artifacts and effective blood signal suppression.

Conclusions: Real-time imaging with low-profile coil designs provides significant enhancements to intravascular MRI.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Aorta / pathology
  • Arteriosclerosis / pathology
  • Artifacts
  • Catheterization / instrumentation*
  • Image Enhancement
  • Image Processing, Computer-Assisted*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Pulsatile Flow
  • Rabbits