J Microbiol Methods. 2002 Jun;50(1):63-73.

Evaluation of redox indicators and the use of digital scanners and spectrophotometer for quantification of microbial growth in microplates.

Author information

1
Microbiology and Tumorbiology Center, Karolinska Institutet, Box 280, 171 77 Stockholm, Sweden. jenny.gabrielson@mtc.ki.se

Abstract

The growth indicators 2,3,5-triphenyltetrazolium chloride (TTC), 2-[4-iodophenyl]-3-[4-dinitrophenyl]-5-phenyltetrazolium chloride (INT), 2,3-bis[2-methoxy-4-nitro-5-sulfophenyl]-2H-tetrazolium-5-carboxanilide inner salt (XTT), 3-[4,5-dimethylthiazol-2-yl]-2,5-diphenyltetrazolium bromide (MTT), and resazurin were tested for their ability to indicate bacterial growth/growth inhibition. Two reading devices were evaluated and compared, a microplate spectrophotometer and a digital flatbed scanner. The bacteria used in the study were cultivated in 96-wells microplates and readings were made after 24 h. The scanned pictures were analysed with a software developed in-house to generate numerical values. It was found that resazurin was difficult to use since it shifts between three colours. MTT and TTC had a high correlation between the spectrophotometer data and the data from the scanned images. The reproducibility was similar for both reading devices. In no case was there a need to resuspend the pellets before reading. Both the XTT and INT showed lower correlations. It is concluded that bacterial growth/growth inhibition can be easily and reproducibly measured from microplate cultivations with a flatbed scanner or with a microplate spectrophotometer.

PMID:
11943359
DOI:
10.1016/s0167-7012(02)00011-8
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Elsevier Science
Loading ...
Support Center