Eur Heart J. 2002 Jan;23(2):124-32.

Predictors and outcomes of stent thrombosis: an intravascular ultrasound registry.

Author information

1
Division of Cardiovascular Medicine, Stanford University School of Medicine, Stanford, California, U.S.A.

Abstract

AIMS:

To investigate whether intravascular ultrasound provides additional information regarding the prediction of stent thrombosis, a retrospective multicentre registry was designed to enrol patients with stent thrombosis following stent deployment under ultrasound guidance.

METHODS AND RESULTS:

A total of 53 patients were enrolled (mean age 61+/-9 years) with stable angina (43%), unstable angina (36%), and post-infarct angina (21%) who underwent intracoronary stenting. The majority had balloon angioplasty alone prior to stenting (94%) with 6% also undergoing rotational atherectomy. The indication for stenting was elective (53%), suboptimal result (32%) and bailout (15%). There were 1.6+/-0.8 stents/artery with 87% undergoing high-pressure dilatation (> or =14 atmospheres). The minimum stent area was 7.7+/-2.8 mm(2)with a mean stent expansion of 81.5+/-21.9%. Overall, 94% of cases demonstrated one abnormal ultrasound finding (stent under-expansion, malapposition, inflow/outflow disease, dissection, or thrombus). Angiography demonstrated an abnormality in only 32% of cases (chi-square=30.0, P<0.001). Stent thrombosis occurred at 132+/-125 h after deployment. Myocardial infarction occurred in 67% and there was an overall mortality of 15%.

CONCLUSION:

On comparison with angiography, the vast majority of stents associated with subsequent thrombosis have at least one abnormal feature by intravascular ultrasound at the time of stent deployment.

Comment in

PMID:
11785994
DOI:
10.1053/euhj.2001.2707