Retinal thickness measurements from optical coherence tomography using a Markov boundary model.

Author information

1
Biomedical Engineering Program and Signal Analysis & Machine Perception Laboratory, Department of Electrical Engineering, The Ohio State University, Columbus 43210-1272, USA.

Abstract

We present a system for detecting retinal boundaries in optical coherence tomography (OCT) B-scans. OCT is a relatively new imaging modality giving cross-sectional images that are qualitatively similar to ultrasound. However, the axial resolution with OCT is much higher, on the order of 10 microm. Objective, quantitative measures of retinal thickness may be made from OCT images. Knowledge of retinal thickness is important in the evaluation and treatment of many ocular diseases. The boundary-detection system presented here uses a one-dimensional edge-detection kernel to yield edge primitives. These edge primitives are rated, selected, and organized to form a coherent boundary structure by use of a Markov model of retinal boundaries as detected by OCT. Qualitatively, the boundaries detected by the automated system generally agreed extremely well with the true retinal structure for the vast majority of OCT images. Only one of the 1450 evaluation images caused the algorithm to fail. A quantitative evaluation of the retinal boundaries was performed as well, using the clinical application of automatic retinal thickness determination. Retinal thickness measurements derived from the algorithm's results were compared with thickness measurements from manually corrected boundaries for 1450 test images. The algorithm's thickness measurements over a 1-mm region near the fovea differed from the corrected thickness measurements by less than 10 microm for 74% of the images and by less than 25 microm (10% of normal retinal thickness) for 98.4% of the images. These errors are near the machine's resolution limit and still well below clinical significance. Current, standard clinical practice involves a qualitative, visual assessment of retinal thickness. A robust, quantitatively accurate system such as ours can be expected to improve patient care.

PMID:
11585207
DOI:
10.1109/42.952728
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for IEEE Engineering in Medicine and Biology Society
    Loading ...
    Support Center