Format

Send to

Choose Destination
See comment in PubMed Commons below
Radiology. 2000 Jun;215(3):904-9.

Autocorrection in MR imaging: adaptive motion correction without navigator echoes.

Author information

1
Department of Diagnostic Radiology, Mayo Clinic and Foundation, Rochester, MN 55905, USA. manduca@mayo.edu

Abstract

A technique for automatic retrospective correction of motion artifacts on magnetic resonance (MR) images was developed that uses only the raw (complex) data from the MR imager and requires no knowledge of patient motion during the acquisition. The algorithm was tested on coronal images of the rotator cuff in a series of 144 patients, and the improvements in image quality were similar to those achieved with navigator echoes. The results demonstrate that autocorrection can significantly reduce motion artifacts in a technically demanding MR imaging application.

[Indexed for MEDLINE]