Am J Respir Crit Care Med. 2000 Apr;161(4 Pt 1):1161-6.

Clinical evaluation of a computer-controlled pressure support mode.

Author information

1
Institut National de la Santé et de la Recherche Médicale (INSERM) U 492, Département de Physiologie, and Service de Réanimation Médicale, Hôpital Henri Mondor, AP-HP, Université Paris 12, Créteil, France.

Abstract

We have designed a computerized system providing closed-loop control of the level of pressure support ventilation (PSV). The system sets itself at the lowest level of PSV that maintains respiratory rate (RR), tidal volume (VT), and end-tidal CO(2) pressure (PET(CO(2))) within predetermined ranges defining acceptable ventilation (i.e., 12 < RR < 28 cycles/min, VT > 300 ml [> 250 if weight < 55 kg], and PET(CO(2)) < 55 mm Hg [< 65 mm Hg if chronic CO(2) retention]). Ten patients received computer-controlled (automatic) PSV and physician-controlled (standard) PSV, in random order, during 24 h for each mode. An estimation of occlusion pressure (P(0.1)) was recorded continuously. The average time spent with acceptable ventilation as previously defined was 66 +/- 24% of the total ventilation time with standard PSV versus 93 +/- 8% with automatic PSV (p < 0.05), whereas the level of PSV was similar during the two periods (17 +/- 4 cm H(2)O versus 19 +/- 6 cm H(2)O). The time spent with an estimated P(0.1) above 4 cm H(2)O was 34 +/- 35% of the standard PSV time versus only 11 +/- 17% of the automatic PSV time (p < 0.01). Automatic PSV increased the time spent within desired ventilation parameter ranges and apparently reduced periods of excessive workload.

PMID:
10764306
PMCID:
PMC2737132
DOI:
10.1164/ajrccm.161.4.9904064
[Indexed for MEDLINE]
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Atypon Icon for PubMed Central
    Loading ...
    Support Center