Format

Send to

Choose Destination
Genomics. 1999 Dec 15;62(3):369-76.

Denaturing high-performance liquid chromatography detects reliably BRCA1 and BRCA2 mutations.

Author information

1
Division of Senology, University of Vienna, Vienna, 1090, Austria.

Abstract

Denaturing high-performance liquid chromatography (DHPLC) is a recently developed method of comparative sequencing based upon heteroduplex detection. To assess the reliability of this method, 180 different mutations (54 deletions, 12 insertions, and 117 single base substitutions) in BRCA1 and BRCA2 were tested. Second, 25 index individuals with complete DHPLC analysis of BRCA1 were reanalyzed by dye-terminator sequencing. Third, 41 index individuals were analyzed concomitantly by both DGGE and DHPLC. Of the 180 different BRCA1 and BRCA2 mutations, 179 showed heterozygous DHPLC elution profiles. Dye-terminator sequencing of the entire BRCA1 gene, including 5592 bp of coding sequence and 5206 bp of flanking noncoding sequence, in 25 index individuals did not reveal additional variants missed by DHPLC. The concomitant analysis of 41 index cases showed that 4 probably disease-associated mutations were identified by DHPLC while only 3 of those 4 sites were detected by denaturing gradient gel electrophoresis. We conclude that DHPLC is a sensitive and cost-effective method for the screening of BRCA1 and BRCA2.

PMID:
10644434
DOI:
10.1006/geno.1999.6026
[Indexed for MEDLINE]