Am Heart J. 1999 May;137(5):958-66.

Real-time, three-dimensional echocardiography: feasibility of dynamic right ventricular volume measurement with saline contrast.

Author information

1
Division of Medicine and Biomedical Engineering, The Center for Emerging Cardiovascular Technologies, Duke University, Durham, NC, USA.

Abstract

BACKGROUND:

The asymmetry and complex shape of the right ventricle have made it difficult to determine right ventricular (RV) volume with 2-dimensional echocardiography. Three-dimensional cardiac imaging improves visualization of cardiac anatomy but is also complex and time consuming. A newly developed volumetric scanning system holds promise of obviating past limitations.

METHODS:

Real-time, transthoracic 3-dimensional echocardiographic images of the right ventricle were obtained with a high-speed volumetric ultrasound system that uses a 16:1 parallel processing schema from a 2.5 MHz matrix phased-array scanner to interrogate an entire pyramidal volume in real time. The instrumentation was used to measure RV volume in 8 excised canine hearts; dynamic real-time 3-dimensional images were also obtained from 14 normal subjects.

RESULTS:

Three-dimensional images were obtained in vitro and in vivo during intravenous hand-agitated saline injection to determine RV volumes. The RV volumes by real-time 3-dimensional echocardiography are well correlated with those of drained in vitro (y = 1.26x - 9.92, r = 0.97, P <.0001, standard error of the estimate = 3.26 mL). For human subjects, the end-diastolic and end-systolic RV volumes were calculated by tracing serial cross-sectional, inclined C scans; functional data were validated by comparing the scans with conventional 2-dimensional echocardiographic indexes of left ventricular stroke volume.

CONCLUSIONS:

These data indicate that RV volume measurements of excised heart by real-time 3-dimensional echocardiography are accurate and that beat-to-beat RV quantitative measurement applying this imaging method is possible. The new application of real-time 3-dimensional echocardiography presents the opportunity to develop new descriptors of cardiac performance.

PMID:
10220647
DOI:
10.1016/s0002-8703(99)70422-9
[Indexed for MEDLINE]