[Application of continuous intra-arterial blood gas monitoring system "Paratrend 7" for pulmonary lavage of a patient with alveolar proteinosis].

[Article in Japanese]

Author information

1
Department of Anesthesia, Kanagawa Cardiovascular and Respiratory Center, Yokohama.

Abstract

We applied a continuous intra-arterial blood gas monitoring system (Paratrend 7) to a patient with pulmonary alveolar proteinosis during pulmonary lavage. Lavage was performed under general anesthesia with one lung ventilation. We inserted the sensor of Patatrend 7 through a 20 G catheter into the radial artery, and monitored pH, PaCO2 and PaO2 continuously throughout the procedure. SpO2 and EtCO2 were also monitored. Saline 1000-1500 ml was instilled and drained repeatedly by volume limited methods. PaO2 values by Paratrend 7 increased during instillation and decreased during drainage of the irrigating fluid. In contrast, PaCO2 value by Paratrend 7 decreased slightly during instillation and increased during drainage. The change of SpO2 was almost the same as that by Paratrend 7, but the response time of pulse oxymetry was a little quicker than Paratrend 7. During the lavage procedure, respiratory and circulatory condition changed very rapidly, and it is necessary to monitor blood gas change intensively. Paratrend 7 is useful as a perioperative monitoring system, but pulse oxymetry might be sufficient during pulmonary lavage considering its cost.