Radiat Prot Dosimetry. 2016 Sep;170(1-4):346-9. doi: 10.1093/rpd/ncw079. Epub 2016 Apr 7.

ENVIRONMENTAL DOSIMETRY WITH THE BeOSL PERSONAL DOSEMETER.

Author information

1
Radiation Physics Group, Technische Universität Dresden, Institute for Nuclear and Particle Physics, Dresden D-01062, Germany jahn@asp.tu-dresden.de.
2
Radiation Physics Group, Technische Universität Dresden, Institute for Nuclear and Particle Physics, Dresden D-01062, Germany.

Abstract

In the course of this work, a dosemeter for measurement of ambient dose equivalent H*(10) based on the BeOSL dosimetry system was designed. Calculations of the energy response of the two-element BeOSL personal dosemeter in terms of H*(10) showed that the response does not fulfil the requirements of IEC 62387 Especially, the response for photon energies of 60-100 keV is too low. It is possible to correct this under-response using a BeOSL two-element dosemeter, a linear algorithm and an additional cover for the dosemeter. The Hp(0.07) element is shielded with lead, and with the help of a linear algorithm, corrections of the energy response are possible. So, the national requirements for an H*(10) area dosemeter in Germany can be fulfilled. The requirements of IEC 62387 are fulfilled for angles of radiation incidence from 0° to 60° for a rotation around the vertical and horizontal axis through the dosemeter, whereas for angles of incidence from 75° to 90°, the requirements are only fulfilled for rotation around the vertical axis through both elements.

PMID:
27060113
DOI:
10.1093/rpd/ncw079
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Silverchair Information Systems
Loading ...
Support Center