Radiat Prot Dosimetry. 2015 Jun;164(4):587-90. doi: 10.1093/rpd/ncv316. Epub 2015 May 15.

Calibration system for radon EEC measurements.

Author information

1
Ural Federal University, Ekaterinburg, Russia.
2
Ural Federal University, Ekaterinburg, Russia Institute of Industrial Ecology UB RAS, Ekaterinburg, Russia.
3
Ural Federal University, Ekaterinburg, Russia Institute of Industrial Ecology UB RAS, Ekaterinburg, Russia michael@ecko.uran.ru.

Abstract

The measurement of radon equivalent equilibrium concentration (EECRn) is very simple and quick technique for the estimation of radon progeny level in dwellings or working places. The most typical methods of EECRn measurements are alpha radiometry or alpha spectrometry. In such technique, the influence of alpha particle absorption in filters and filter effectiveness should be taken into account. In the authors' work, it is demonstrated that more precise and less complicated calibration of EECRn-measuring equipment can be conducted by the use of the gamma spectrometer as a reference measuring device. It was demonstrated that for this calibration technique systematic error does not exceed 3 %. The random error of (214)Bi activity measurements is in the range 3-6 %. In general, both these errors can be decreased. The measurements of EECRn by gamma spectrometry and improved alpha radiometry are in good agreement, but the systematic shift between average values can be observed.

PMID:
25979737
DOI:
10.1093/rpd/ncv316
[Indexed for MEDLINE]