This book is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) ( http://creativecommons.org/licenses/by-nc-nd/4.0/ ), which permits others to distribute the work, provided that the article is not altered or used commercially. You are not required to obtain permission to distribute this article, provided that you credit the author and journal.
NCBI Bookshelf. A service of the National Library of Medicine, National Institutes of Health.
StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2025 Jan-.

StatPearls [Internet].
Show detailsIntroduction
Electrochemiluminescence combines electrochemical reactions and luminescence, converting electrical energy to light.[1] Electrochemiluminescence differs from chemiluminescence; in electrochemiluminescence, the reactive species that produce the chemiluminescent reaction are electrochemically generated from stable precursors at the surface of an electrode.[2]
Luminophores are substances that emit light. In electrochemiluminescence, luminophores attain a high-energy state induced by electron transfer at the electrode surface through an oxidation-reduction reaction. The excited luminophores emit light as photons while returning to the ground state.[1] Luminophores can be used as labels for biomolecules; the biomolecules can be detected and quantified by measuring the amount of light emitted.[3]
Electrochemiluminescence is an important diagnostic technique known for its versatility and numerous advantages. The applications of electrochemiluminescence include detecting, separating, and quantifying various intracellular and extracellular biomolecules, including proteins, enzymes, hormones, metabolites, and nucleic acids.[4] Electrochemiluminescence is also used to visualize cells, study the functions of various intramembrane and transmembrane proteins, detect nucleic acids of interest, and assay drugs.[3]
Electrochemiluminescence systems are classified into 2 types—ion annihilation or co-reactant systems.
Ion Annihilation Electrochemiluminescence System
In ion annihilation systems, a pulsed potential applied to the electrode generates radical cations and anions of the luminophore.[3][5] The electron transfer between anions and cations results in an excited cation. The subsequent decay of this excited cation to the ground state results in the emission of light. Ion annihilation systems typically use organic compounds dissolved in organic solvents, donor-acceptor conjugated molecules.[6] These organic compounds are typically poor candidates for biomolecular assay labels. Ion annihilation also generates highly reactive intermediates unsuitable for routine assays.[7]
Co-reactant Electrochemiluminescence S ystem
The majority of electrochemiluminescence systems currently in use are co-reactant systems.[6] These systems use a high-efficiency co-reactant added to the luminophore with one-directional potential scanning. Oxidation or reduction of both species at the electrode generates radicals. Intermediates from the co-reactant decompose, forming a robust species that reacts with the luminophore, producing excited states and emitting light. Co-reactant systems are used for biomolecular assays due to the solubility of the co-reactant in the surrounding medium, low reduction-oxidation potential, and stability.[7] Ruthenium metal ions and luminol derivatives are the most commonly used lumiphores in co-reactant electrochemiluminescence systems.[8] The co-reactant commonly used with ruthenium metal ions is tripropylamine. Other widely used co-reactants include 2-(dibutylamino)ethanol, peroxydisulfate, and hydrogen peroxide.[9]
Most reported electrochemiluminescence applications for immunoassay or genetic analysis use tris(2,2′-bipyridyl)ruthenium as a label and tripropylamine as a co-reactant.[10] These systems are highly efficient, as the ruthenium compound is stable, highly soluble in polar and nonpolar solvents, and exhibits strong luminescence. Tripropylamine undergoes oxidation with potential application, forming a tripropylamine radical cation and a tripropylamine radical. These radicals generate excited bipyridyl-ruthenium, which emits orange-spectrum light at 600 to 640 nm as it relaxes to the ground state. The luminophore is regenerated after emission.[11]
Luminol is an organic luminophore commonly used for cell imaging.[3] Upon oxidation, it forms a diazaquinone intermediate, which further oxidizes to 3-aminophthalate in the presence of hydrogen peroxide, emitting blue light. Hydrogen peroxide, generated in biological processes, is often detected alongside luminol. Reactive oxygen species can enhance luminol electrochemiluminescence emission. Luminol is irreversibly oxidized and requires alkaline conditions, limiting cellular analysis applications. However, it operates at a lower anodic potential compared to bipyridyl-ruthenium, providing advantages for imaging living cells.[12]
Basic Instrumentation for Electrochemiluminescence
The basic instrumentation setup in an electrochemiluminescence includes an electrochemical cell, a detector, a signal amplification system, and a reagent and sample delivery system.[7] The electrochemical cell houses the working electrode, typically made of carbon or gold, which serves as a site of the electrochemiluminescence reaction. A reference electrode is also present to maintain a stable potential for measurement. A photomultiplier tube or a photodiode is commonly used to detect the light emitted during the electrochemiluminescence reaction. These detectors are highly sensitive to low light levels and convert the photons into electrical signals. The electrical signals generated by the light detection system are weak and must be amplified and processed for accurate measurement. Amplification circuits, such as transimpedance amplifiers, can boost signal strength, and signal processing units can filter and digitize the signal for analysis. In an electrochemiluminescence assay, reagents and samples are delivered to the electrochemical cell using a syringe pump or a microfluidic system, which precisely administers the required volumes at specific time points.
Advantages and Limitations of Electrochemiluminescence
Electrochemiluminescence systems have several advantages. The luminophores used in electrochemiluminescence are small, stable substances that can label a wide range of molecules and haptens without cross-reaction. There is minimal background interference in electrochemiluminescence because the luminophore has the inherent capacity to emit light, and no additional light source is required. The technique is susceptible due to multiple excitation cycles and enables detection at very low limits, as low as 200 fmol/L. In addition, electrochemiluminescence provides improved reagent stability.[4][7]
Electrochemiluminescence is susceptible to light leaks and background luminescence from reagents. The high sensitivity offered by electrochemiluminescence requires pure reagents and solvents. In addition, high-intensity light emission may lead to pulse pile-up, resulting in underestimating light emission.[13][14]
Specimen Requirements and Procedure
Electrochemiluminescence assays in the clinical setting require careful sample handling to ensure accurate and reliable results. The process involves the following steps:
Sample Collection
The first step in an electrochemiluminescence assay is collecting a clinical sample, such as blood, urine, or serum, from the patient. Proper collection techniques and appropriate sample containers are essential to maintain sample integrity and prevent contamination. A 200- to 400-µL sample is required for any electrochemiluminescence assay.[15] Adequate samples should be available for single-sample and entire-run repeat analyses in case of quality control (QC) failure.
Sample Preparation
Samples containing precipitates should be centrifuged before testing. Heat-inactivated samples should not be used. Samples should be stored frozen upon receipt unless they are to be analyzed immediately and should be assayed without delay after thawing. Multiple freeze-thaw cycles should be avoided; samples may lose their integrity.[16] For the nucleic acid assay, specimens are processed to extract the DNA or RNA of interest using appropriate isolation methods. The purity and concentration of the nucleic acids are determined using spectrophotometry or fluorometry.[17]
Sample Dilution
In some cases, the concentration of the target analyte in the clinical sample may be too high or too low for accurate measurement. In such instances, appropriate sample dilutions may be necessary to bring the analyte concentration within the assay's linear range.[15] Dilution protocols should be followed precisely to ensure accurate results.
Reagent Mixing
Electrochemiluminescence assays often involve adding specific reagents to the sample to initiate the electrochemiluminescence reaction. Proper mixing of the sample and reagents is essential to ensure uniform distribution and complete reaction. Depending on the assay requirements, mixing can be achieved by vortexing, pipetting, or using automated mixing devices.
Incubation
Some electrochemiluminescence assays may require a specific incubation period to allow the electrochemiluminescence reaction to proceed wholly. During this time, the sample and the added reagents are kept at a controlled temperature to promote the reaction. The duration and temperature of the incubation should be precise to achieve optimal assay performance.
Calibration Standards and Controls
Calibration standards and controls are typically included to validate the accuracy and precision of the electrochemiluminescence assay. Calibration standards with known target analyte concentrations are used to generate a calibration curve, which relates the electrochemiluminescence signal to the analyte concentration.[16] Control samples with known values monitor the assay performance and detect potential issues.
Diagnostic Tests
Electrochemiluminescence methods are used in the following specialized clinical assays: [4][17]
Hormones
- thyroid-stimulating hormone (TSH), thyroxine (T4), and triiodothyronine (T3)
- Progesterone
- Testosterone
- Follicle-stimulating hormone (FSH)
- Luteinizing hormone (LH)
- Prolactin
Tumor Markers
- Carcinoembryonic antigen
- Carbohydrate antigen-125, 19.9, and 15.3
- Prostrate-specific antigen
- Beta-human chorionic gonadotropin
Therapeutic Drug Monitoring
- Carbamazepine
- Digoxin
- Phenytoin
- Vancomycin
Vitamins
- Vitamin B12
- Folic acid
- Vitamin D
Proteins
- Ferritin
- D-dimer
- Brain natriuretic polypeptide
- Atrial natriuretic polypeptide
- Insulin
- C-peptide
Inflammatory Markers
- C-reactive protein
- Interleukins
- Tumor necrosis factor
- Growth factors
- Complements
Antibodies
- Anticardiolipin antibodies
- Anti-thyroid peroxidase antibody
- Anti-thyroglobulin antibody
Infectious Diseases
- Herpes simplex virus antibody
- Rubella virus antibody
- Toxoplasma circulating antibody
- Hepatitis E antigen and antibody
- Coxsackievirus antibody
- SARS-Co-V antigen and antibody
Testing Procedures
Electrochemiluminescence Immunoassay
The electrochemiluminescence immunoassay (ECLIA) relies on the interaction between an antibody and its corresponding antigen. The ECLIA uses specialized reagents, including a capture antibody against the antigen, typically a biomolecule, and a labeled antibody to detect the interaction. The capture antibody is immobilized on a solid support, such as a microplate or magnetic bead. In contrast, the labeled antibody is conjugated with a luminescent marker and an electrochemically active molecule.[18] Many studies have proven ECLIA superior to conventional radioimmunoassay and enzyme-linked immunosorbent assays (ELISA); ECLIA is highly sensitive and uses neither a radioisotope nor enzymes with limited stability.[19][20]
ECLIA uses 3 basic interaction principles to detect analytes—direct, competitive, and sandwich. In a direct ECLIA system, a single antibody is immobilized on the electrode. The electrode can capture the target analyte (Ag) within the specimen and act as an electrochemiluminescent probe. The Ag competes with the co-reactant for binding sites on the antibody; an increase in bound Ag decreases the electrochemiluminescence signal. The quantifiable decrease in the electrochemiluminescence signal directly correlates to the concentration of the Ag.[21] In a competitive interaction, an antigen analog is labeled with a luminophore, and the antibody is immobilized on the surface of the electrode.[21] The Ag in the analyte competes with the analog to bind to the antibody; the concentration of labels in the system decreases. The concentration of the Ag is inversely proportional to the measured electrochemiluminescence signal. The competitive ECLIA is used for smaller Ag such as thyroid hormones, cortisol, and testosterone.[22] Lastly, a sandwich interaction uses 2 different antibodies. The primary antibody is typically immobilized on the electrode surface as a capture probe for the target antigen. The secondary antibody also targets the antigen and is labeled with a luminophore. The target antigen binds to form an immunocomplex with the capture probe and the electrochemiluminescence probe, and the measured electrochemiluminescence signal intensity is proportional to the concentration of the target antigen.[23] The sandwich ECLIA is used for larger analytes such as TSH, FSH, and LH.[22]
Bioconjugation
Bioconjugation involves linking the luminophore with biological samples such as antigens or antibodies. There are various linking groups suitable for ruthenium complexes to conjugate with biological samples, such as phosphoramidite for conjugation with oligonucleotides, N-hydroxysuccinimide ester for linking with amines on proteins, hydrazide for carbohydrates, maleimide for thiols, and amines as linking groups for reactions with carboxylic acids on proteins.[24][25] Surface modification with biotin-streptavidin is performed to capture antibodies or antigens on the surface support. The binding between streptavidin and biotin provides a strong and specific interaction immobilizing the antibody on the surface. The subsequent formation of a sandwich complex and the generation of the electrochemiluminescence signal upon electrochemical stimulation enable the sensitive detection of the target analyte in the ECLIA.[26]
Nucleic Acid Detection
Specific DNA probes or primers are designed to target the desired DNA or RNA sequences. Labeled with luminophores, these probes can hybridize with the complementary target DNA, bringing the electrochemiluminescence-active molecule close to the electrode surface and generating luminescence.[27] These assays are used in genetic analysis, mutation detection, pathogen identification, and other molecular biology and diagnostics fields.
Cellular Imaging
Living cells are typically labeled with electrochemiluminescence-active probes or antibodies specific to cellular markers. The electrochemiluminescence-based detection allows visualization and analysis of cellular processes, protein-protein interactions, and intracellular signaling pathways.[28]
Interfering Factors
Endogenous Factors
Most commercially available ECLIA-based immunoassays are generally unaffected by endogenous interfering substances.[29] Hemoglobin, bilirubin, and lipids are the most common endogenous interfering substances. However, serum bilirubin up to 64 mg/dL, lipemia up to 1900 mg/dL, and serum hemoglobin up to 1 g/dL do not interfere with ECLIA-based immunoassays.
Historically, the cross-reaction of endogenous substances, such as LH interfering with human chorionic gonadotropin, was problematic. However, with the advent of more specific antibodies, cross-reaction is minimal. Still, cross-reactivity can be a problem during the detection of drugs and their metabolites.[30]
Heterophilic antibodies such as rheumatoid factor, anti-thyroid antibodies, and anti-animal antibodies may interfere through noncompetitive mechanisms, resulting in falsely high or low values; the incidence varies from 0.5% to 6%.[31][32] Many cases of antibodies against streptavidin and ruthenium complexes have been reported.[33]
Exogenous Factors
ECLIA-based immunoassays should be avoided in patients taking > 5 mg/day of therapeutic biotin.[34] Temporary discontinuation of biotin is advised before performing the assay. When sodium citrate is used as an anticoagulant, the results must be corrected by more than 10%.[35] Samples or assay reagents contaminated with substances might interfere with the measurement of the label.
High-Dose Hook Effect
The high-dose hook effect, or prozone phenomenon, is particular to the sandwich immunoassay. This effect is characterized by a false-negative or underestimated result at high target analyte concentrations. In the high-dose hook effect, an extremely high concentration of the target analyte can exceed the binding capacity of the capture and labeled antibodies.[24] As a result, the excess analyte saturates both the captured and labeled antibody-binding sites, preventing the formation of the sandwich complex. This saturation leads to decreased signal or signal intensity, resulting in a false-negative or underestimated result. Dilution of the sample is a commonly used strategy to mitigate this effect.[36]
Results, Reporting, and Critical Findings
Electrochemiluminescence provides highly sensitive and specific measurements, enabling the detection and quantification of analytes with exceptional precision. Signals are displayed as relative light units (RLU), representing the amount of light emitted.[4] The RLU is converted into concentration units by a calibration curve plotted using various analyte concentrations and signals obtained.
When reporting results, it is essential to provide accurate and comprehensive information, including the measured analyte concentration, units of measurement, and any necessary interpretation or reference ranges.[37] Critical findings, defined as values outside established thresholds or diagnostic cutoffs, require immediate attention and notification to the healthcare team. These findings may indicate significant health risks or the need for urgent medical interventions.[18]
Clinical Significance
Endocrinopathies
The high sensitivity, wide dynamic range, multiplexing capability, precision, and high throughput analysis of electrochemiluminescence make it a valuable tool for detecting and monitoring hormonal disorders. With fully automated analyzers and commercially available kits, ECLIA-based systems are superior to conventional radioimmunoassay and ELISA. These assays are used when diagnosing and managing conditions such as hypothyroidism and hyperthyroidism to quantify circulating levels of TSH, T4, and T3.[38] Insulin and C-peptide measurements assist in diagnosing and managing type 1 and type 2 diabetes. Electrochemiluminescence is used to detect adrenal disorders, such as adrenal insufficiency and the hypercortisolism of Cushing syndrome. Reproductive hormone assays, including LH, FSH, estrogen, progesterone, and testosterone, aid in the evaluation and management of polycystic ovary syndrome, infertility, and hormonal imbalances.[39][40]
Detection and Monitoring of Neoplasms
Prostate-specific antigen is a widely used biomarker for prostate cancer. Tumor antigens such as carcinoembryonic antigen and carbohydrate antigen 19-9 assist in diagnosing, prognosis, and monitoring disease progression in colorectal and pancreatic adenocarcinomas.[41]
Infectious Processes
Electrochemiluminescence-based assays detect HIV antibodies or viral antigens, such as p24, hepatitis B surface antigen, hepatitis C virus antibodies, or viral antigens.[42] Electrochemiluminescence-based assays aid in detecting Mycobacterium tuberculosis infections by measuring antibodies or antigens specific to the bacillus, such as interferon-gamma release assays or M tuberculosis antigen 85 complexes.[43] Antigen-based electrochemiluminescence assays are also available to diagnose coronavirus infections.[44]
Biosensors and Point-of-Care Testing
Biosensors are miniaturized analytical systems incorporating a biological recognition element that selectively interacts with a target analyte. This recognition element can be an enzyme, antibody, nucleic acid, or other biomolecule that exhibits a specific binding affinity toward the target analyte. Recent innovations, such as using Ru-loaded silica and gold nanoparticles, integrating the microfluidic system, and using screen print nanoelectrodes, have facilitated the integration of biosensors as point-of-care devices.[8]
Electrochemiluminescence-based point-of-care-testing biosensors aid in detecting biomarkers such as cardiac troponin I and C-reactive protein, enabling early diagnosis of acute myocardial infarction and monitoring of postoperative recovery.[36] Integrating electrochemiluminescence with mobile devices and highly integrated systems ensures portability, simplicity, and accessibility for effective disease control and surveillance in public health. Electrochemiluminescence biosensors based on DNA hybridization principles are used for genetic analysis and genotyping. Aptamers, synthetic single-stranded DNA or RNA sequences with high affinity and specificity for target molecules, are recognition elements in electrochemiluminescence biosensors.[45]
Quality Control and Lab Safety
QC and laboratory safety are critical when using the electrochemiluminescence system for clinical assays. QC procedures include regular calibration and verification using multi-level lyophilized control materials to ensure system precision and accuracy.[46] For non-waived tests, laboratory regulations require, at the minimum, analysis of at least 2 levels of QC materials once every 24 hours. If necessary, laboratories can assay QC samples more frequently to ensure accurate results.[47]
QC samples should be assayed after calibration or maintenance of an analyzer to verify the correct method performance.[48] To minimize QC for tests where manufacturers' recommendations are less frequent than those required by the regulatory agency, such as once per month, laboratories can develop an individualized QC plan. This plan involves performing a risk assessment of potential sources of error in all testing phases and implementing a QC plan to reduce the likelihood of errors.[49]
The design of a QC plan must consider the analytical performance capability of a measurement procedure and the risk of harm to a patient if an erroneous laboratory test result is used for a clinical care decision. An erroneous laboratory test result is a hazardous condition that may or may not cause harm to a patient, depending on what action or inaction a clinical care provider takes based on the erroneous result.[50]
The acceptable range and rules for interpreting QC results are based on the probability of detecting a significant analytical error condition with an acceptably low false alert rate.[48] The desired process control performance characteristics must be established for each measurement before selecting the appropriate QC rules.[50] Westgard multi-rules are typically used to evaluate the QC runs. If QC results are out of range, the system should be investigated to determine the cause of the problem, and analysis should be paused until the issue is resolved.[51]
Changes in reagent lots can unexpectedly impact QC results, necessitating careful reagent lot crossover evaluation of QC target values. As the matrix-related interaction between a QC material and a reagent can change with a different reagent lot, QC results may not be a reliable indicator of a measurement procedure's performance for patient samples after a reagent lot change.[52] Clinical patient samples should be used to verify the consistency of results between old and new lots of reagents because of the unpredictability of a matrix-related bias being present for QC materials.[53]
Participation in external QC or proficiency testing programs is mandated by the Clinical Laboratory Improvement Amendments and overseen by the Centers for Medicare and Medicaid Services (CMS).[54] These programs ensure a lab's accuracy compared to others performing similar assays. Required participation and scored CMS and voluntary accreditation organizations monitor results. The proficiency testing plan should be included in the quality assessment plan and the laboratory's overall quality program.[55]
Simultaneously, maintaining lab safety is crucial by performing daily maintenance, preventive maintenance, proper disposal of waste, using personal protective equipment, following proper safety protocols, handling and storing hazardous reagents, and adhering to safety guidelines. By prioritizing QC and lab safety, laboratories can deliver precise and dependable clinical assay results while safeguarding the well-being of staff and patients.[56]
Enhancing Healthcare Team Outcomes
Optimal team performance is essential for applying an electrochemiluminescence system in patient-centered care. Healthcare workers handling electrochemiluminescence systems should possess expertise in instrumentation, sample handling, report interpretation, and troubleshooting, which can be developed through continuous training and skill development.
When implementing electrochemiluminescence systems, factors such as sensitivity, specificity, and cost-effectiveness should be carefully evaluated for each clinical setting, with collaborative input from healthcare practitioners, laboratory professionals, and technicians. Clear and concise communication regarding patient history, test requirements, and interpretation of electrochemiluminescence results promotes accurate diagnosis, appropriate treatment decisions, and coordinated care.
References
- 1.
- Brown K, Blake RS, Dennany L. Electrochemiluminescence within veterinary Science: A review. Bioelectrochemistry. 2022 Aug;146:108156. [PubMed: 35598500]
- 2.
- Wang J, Wang MR. [Comparative analysis of electro-chemiluminescence immunoassay and chemiluminescent microparticle immunoassay abilities to quantitatively assess hepatitis B surface antigen]. Zhonghua Gan Zang Bing Za Zhi. 2013 Mar;21(3):192-5. [PubMed: 23967740]
- 3.
- Miao W. Electrogenerated chemiluminescence and its biorelated applications. Chem Rev. 2008 Jul;108(7):2506-53. [PubMed: 18505298]
- 4.
- Yoo SM, Jeon YM, Heo SY. Electrochemiluminescence Systems for the Detection of Biomarkers: Strategical and Technological Advances. Biosensors (Basel). 2022 Sep 07;12(9) [PMC free article: PMC9496345] [PubMed: 36140123]
- 5.
- Al-Kutubi H, Voci S, Rassaei L, Sojic N, Mathwig K. Enhanced annihilation electrochemiluminescence by nanofluidic confinement. Chem Sci. 2018 Dec 28;9(48):8946-8950. [PMC free article: PMC6301198] [PubMed: 30647886]
- 6.
- Huang P, Zou X, Xu Z, Lan Y, Chen L, Zhang B, Niu L. Studies on Annihilation and Coreactant Electrochemiluminescence of Thermally Activated Delayed Fluorescent Molecules in Organic Medium. Molecules. 2022 Nov 02;27(21) [PMC free article: PMC9658960] [PubMed: 36364282]
- 7.
- Zhao W, Chen HY, Xu JJ. Electrogenerated chemiluminescence detection of single entities. Chem Sci. 2021 Mar 18;12(16):5720-5736. [PMC free article: PMC8179668] [PubMed: 34168801]
- 8.
- Martínez-Periñán E, Gutiérrez-Sánchez C, García-Mendiola T, Lorenzo E. Electrochemiluminescence Biosensors Using Screen-Printed Electrodes. Biosensors (Basel). 2020 Sep 09;10(9) [PMC free article: PMC7559215] [PubMed: 32916838]
- 9.
- Liu X, Shi L, Niu W, Li H, Xu G. Environmentally friendly and highly sensitive ruthenium(II) tris(2,2'-bipyridyl) electrochemiluminescent system using 2-(dibutylamino)ethanol as co-reactant. Angew Chem Int Ed Engl. 2007;46(3):421-4. [PubMed: 17146822]
- 10.
- Zu Y, Bard AJ. Electrogenerated chemiluminescence. 66. The role of direct coreactant oxidation in the ruthenium tris(2,2')bipyridyl/tripropylamine system and the effect of halide ions on the emission intensity. Anal Chem. 2000 Jul 15;72(14):3223-32. [PubMed: 10939391]
- 11.
- Miao W, Choi JP, Bard AJ. Electrogenerated chemiluminescence 69: the tris(2,2'-bipyridine)ruthenium(II), (Ru(bpy)3(2+))/tri-n-propylamine (TPrA) system revisited-a new route involving TPrA*+ cation radicals. J Am Chem Soc. 2002 Dec 04;124(48):14478-85. [PubMed: 12452725]
- 12.
- Khan P, Idrees D, Moxley MA, Corbett JA, Ahmad F, von Figura G, Sly WS, Waheed A, Hassan MI. Luminol-based chemiluminescent signals: clinical and non-clinical application and future uses. Appl Biochem Biotechnol. 2014 May;173(2):333-55. [PMC free article: PMC4426882] [PubMed: 24752935]
- 13.
- Hao N, Wang K. Recent development of electrochemiluminescence sensors for food analysis. Anal Bioanal Chem. 2016 Oct;408(25):7035-48. [PubMed: 27086020]
- 14.
- Ravalli A, Voccia D, Palchetti I, Marrazza G. Electrochemical, Electrochemiluminescence, and Photoelectrochemical Aptamer-Based Nanostructured Sensors for Biomarker Analysis. Biosensors (Basel). 2016 Aug 02;6(3) [PMC free article: PMC5039658] [PubMed: 27490578]
- 15.
- Aoyama M, Mano Y. Application of an electrochemiluminescence assay for quantification of E6011, an antifractalkine monoclonal antibody, to pharmacokinetic studies in monkeys and humans. J Clin Lab Anal. 2019 Jan;33(1):e22625. [PMC free article: PMC6430341] [PubMed: 30030862]
- 16.
- Findlay JW, Smith WC, Lee JW, Nordblom GD, Das I, DeSilva BS, Khan MN, Bowsher RR. Validation of immunoassays for bioanalysis: a pharmaceutical industry perspective. J Pharm Biomed Anal. 2000 Jan;21(6):1249-73. [PubMed: 10708409]
- 17.
- Blackburn GF, Shah HP, Kenten JH, Leland J, Kamin RA, Link J, Peterman J, Powell MJ, Shah A, Talley DB. Electrochemiluminescence detection for development of immunoassays and DNA probe assays for clinical diagnostics. Clin Chem. 1991 Sep;37(9):1534-9. [PubMed: 1716534]
- 18.
- Wheeler MJ. Immunoassay techniques. Methods Mol Biol. 2006;324:1-23. [PubMed: 16761369]
- 19.
- Kuhle J, Barro C, Andreasson U, Derfuss T, Lindberg R, Sandelius Å, Liman V, Norgren N, Blennow K, Zetterberg H. Comparison of three analytical platforms for quantification of the neurofilament light chain in blood samples: ELISA, electrochemiluminescence immunoassay and Simoa. Clin Chem Lab Med. 2016 Oct 01;54(10):1655-61. [PubMed: 27071153]
- 20.
- Stockmann W, Bablok W, Luppa P. Analytical performance of Elecsys 2010--a multicentre evaluation. Wien Klin Wochenschr. 1998;110 Suppl 3:10-21. [PubMed: 9677668]
- 21.
- Erler K. Elecsys immunoassay systems using electrochemiluminescence detection. Wien Klin Wochenschr. 1998;110 Suppl 3:5-10. [PubMed: 9677667]
- 22.
- Tabatabaei MS, Ahmed M. Enzyme-Linked Immunosorbent Assay (ELISA). Methods Mol Biol. 2022;2508:115-134. [PubMed: 35737237]
- 23.
- Kitamura K, Matsuda K, Ide M, Tokunaga T, Honda M. A fluorescence sandwich ELISA for detecting soluble and cell-associated human interleukin-2. J Immunol Methods. 1989 Jul 26;121(2):281-8. [PubMed: 2788194]
- 24.
- Rusling JF, Forster RJ. Biosensors Designed for Clinical Applications. Biomedicines. 2021 Jun 22;9(7) [PMC free article: PMC8301448] [PubMed: 34206405]
- 25.
- Asbaghian-Namin H, Karami P, Naghshara H, Gholamin D, Johari-Ahar M. Electrochemiluminescent immunoassay for the determination of CA15-3 and CA72-4 using graphene oxide nanocomposite modified with CdSe quantum dots and Ru(bpy)3 complex. Mikrochim Acta. 2021 Jun 28;188(7):238. [PubMed: 34184115]
- 26.
- Flamme M, Clarke E, Gasser G, Hollenstein M. Applications of Ruthenium Complexes Covalently Linked to Nucleic Acid Derivatives. Molecules. 2018 Jun 22;23(7) [PMC free article: PMC6099586] [PubMed: 29932443]
- 27.
- Gao X, Ren X, Ai Y, Li M, Zhang B, Zou G. Dual-potential encoded electrochemiluminescence for multiplexed gene assay with one luminophore as tag. Biosens Bioelectron. 2023 Sep 15;236:115418. [PubMed: 37279619]
- 28.
- Hiramoto K, Villani E, Iwama T, Komatsu K, Inagi S, Inoue KY, Nashimoto Y, Ino K, Shiku H. Recent Advances in Electrochemiluminescence-Based Systems for Mammalian Cell Analysis. Micromachines (Basel). 2020 May 22;11(5) [PMC free article: PMC7281524] [PubMed: 32456040]
- 29.
- Klee GG. Harmonization and standardization of thyroid function tests. Clin Chem. 2010 Jun;56(6):879-80. [PubMed: 20378766]
- 30.
- Nickoloff EL. Interference in immunoassay. Crit Rev Clin Lab Sci. 1984;21(3):255-67. [PubMed: 6391818]
- 31.
- Zouwail SA, O'Toole AM, Clark PM, Begley JP. Influence of thyroid hormone autoantibodies on 7 thyroid hormone assays. Clin Chem. 2008 May;54(5):927-8. [PubMed: 18443182]
- 32.
- Ellis MJ, Livesey JH. Techniques for identifying heterophile antibody interference are assay specific: study of seven analytes on two automated immunoassay analyzers. Clin Chem. 2005 Mar;51(3):639-41. [PubMed: 15650033]
- 33.
- Ekins R. Immunoassay standardization. Scand J Clin Lab Invest Suppl. 1991;205:33-46. [PubMed: 1947747]
- 34.
- Liu D, Gebreab YB, Hu J, Zhou L, Zhang N, Tong H, Chen B, Wang X. Development and Evaluation of an Anti-Biotin Interference Method in Biotin-Streptavidin Immunoassays. Diagnostics (Basel). 2022 Jul 16;12(7) [PMC free article: PMC9324022] [PubMed: 35885633]
- 35.
- Holmes TH. Generalized mathematical model for immunoassay interference. Autoimmun Rev. 2020 Nov;19(11):102663. [PubMed: 32942040]
- 36.
- Al-Mahdili HA, Jones GR. High-dose hook effect in six automated human chorionic gonadotrophin assays. Ann Clin Biochem. 2010 Jul;47(Pt 4):383-5. [PubMed: 20511370]
- 37.
- Babamiri B, Bahari D, Salimi A. Highly sensitive bioaffinity electrochemiluminescence sensors: Recent advances and future directions. Biosens Bioelectron. 2019 Oct 01;142:111530. [PubMed: 31398687]
- 38.
- Kazerouni F, Amirrasouli H. Performance characteristics of three automated immunoassays for thyroid hormones. Caspian J Intern Med. 2012 Spring;3(2):400-104. [PMC free article: PMC3861902] [PubMed: 24358433]
- 39.
- Cánovas R, Daems E, Campos R, Schellinck S, Madder A, Martins JC, Sobott F, De Wael K. Novel electrochemiluminescent assay for the aptamer-based detection of testosterone. Talanta. 2022 Mar 01;239:123121. [PubMed: 34942485]
- 40.
- Apilux A, Rengpipat S, Suwanjang W, Chailapakul O. Paper-based immunosensor with competitive assay for cortisol detection. J Pharm Biomed Anal. 2020 Jan 30;178:112925. [PubMed: 31669910]
- 41.
- Zhang Y, Liu W, Ge S, Yan M, Wang S, Yu J, Li N, Song X. Multiplexed sandwich immunoassays using flow-injection electrochemiluminescence with designed substrate spatial-resolved technique for detection of tumor markers. Biosens Bioelectron. 2013 Mar 15;41:684-90. [PubMed: 23062558]
- 42.
- Liao Y, Zhou X, Fu Y, Xing D. Linear Ru(bpy)32+-Polymer as a Universal Probe for Sensitive Detection of Biomarkers with Controllable Electrochemiluminescence Signal-Amplifying Ratio. Anal Chem. 2017 Dec 05;89(23):13016-13023. [PubMed: 29092402]
- 43.
- Zhou B, Zhu M, Hao Y, Yang P. Potential-Resolved Electrochemiluminescence for Simultaneous Determination of Triple Latent Tuberculosis Infection Markers. ACS Appl Mater Interfaces. 2017 Sep 13;9(36):30536-30542. [PubMed: 28828860]
- 44.
- Drobysh M, Ramanaviciene A, Viter R, Chen CF, Samukaite-Bubniene U, Ratautaite V, Ramanavicius A. Biosensors for the Determination of SARS-CoV-2 Virus and Diagnosis of COVID-19 Infection. Int J Mol Sci. 2022 Jan 08;23(2) [PMC free article: PMC8776074] [PubMed: 35054850]
- 45.
- Hanif A, Farooq R, Rehman MU, Khan R, Majid S, Ganaie MA. Aptamer based nanobiosensors: Promising healthcare devices. Saudi Pharm J. 2019 Mar;27(3):312-319. [PMC free article: PMC6438676] [PubMed: 30976173]
- 46.
- Ain QU, Younas M, Haroon ZH, Kirmani SI, Munir MU, Anwar M. Fifth Generation Quality Control of Immunoassay Parameters by Six Sigma Value. J Coll Physicians Surg Pak. 2023 Jun;33(6):679-683. [PubMed: 37300265]
- 47.
- Kinns H, Pitkin S, Housley D, Freedman DB. Internal quality control: best practice. J Clin Pathol. 2013 Dec;66(12):1027-32. [PubMed: 24072731]
- 48.
- Kearney E. Internal quality control. Methods Mol Biol. 2013;1065:277-89. [PubMed: 23996371]
- 49.
- Westgard JO. A Total Quality-Control Plan with Right-Sized Statistical Quality-Control. Clin Lab Med. 2017 Mar;37(1):137-150. [PubMed: 28153361]
- 50.
- Westgard JO. Internal quality control: planning and implementation strategies. Ann Clin Biochem. 2003 Nov;40(Pt 6):593-611. [PubMed: 14629798]
- 51.
- Bayat H. Selecting multi-rule quality control procedures based on patient risk. Clin Chem Lab Med. 2017 Oct 26;55(11):1702-1708. [PubMed: 28236626]
- 52.
- Topic E, Nikolac N, Panteghini M, Theodorsson E, Salvagno GL, Miler M, Simundic AM, Infusino I, Nordin G, Westgard S. How to assess the quality of your analytical method? Clin Chem Lab Med. 2015 Oct;53(11):1707-18. [PubMed: 26408611]
- 53.
- Dalenberg DA, Schryver PG, Klee GG. Analytical performance specifications: relating laboratory performance to quality required for intended clinical use. Clin Lab Med. 2013 Mar;33(1):55-73. [PubMed: 23331729]
- 54.
- Badrick T. Integrating quality control and external quality assurance. Clin Biochem. 2021 Sep;95:15-27. [PubMed: 33965412]
- 55.
- Centers for Disease Control and Prevention (CDC) (2) Centers for Medicare & Medicaid Services (CMS), HHS. Medicare, Medicaid, and CLIA programs; laboratory requirements relating to quality systems and certain personnel qualifications. Final rule. Fed Regist. 2003 Jan 24;68(16):3639-714. [PubMed: 12545998]
- 56.
- Lo J. Biological safety in the medical laboratory. Hong Kong Med J. 2015 Jun;21(3):200. [PubMed: 26045068]
Disclosure: Sujatha Mahadevarao Premnath declares no relevant financial relationships with ineligible companies.
Disclosure: Muhammad Zubair declares no relevant financial relationships with ineligible companies.
- Review Electrogenerated chemiluminescence.[Annu Rev Anal Chem (Palo Alto ...]Review Electrogenerated chemiluminescence.Forster RJ, Bertoncello P, Keyes TE. Annu Rev Anal Chem (Palo Alto Calif). 2009; 2:359-85.
- Review Ion cyclotron resonance spectroscopy. Cyclotron double resonance provides a new technique for the study of ion-molecule reaction mechanisms.[Science. 1968]Review Ion cyclotron resonance spectroscopy. Cyclotron double resonance provides a new technique for the study of ion-molecule reaction mechanisms.Baldeschwieler JD. Science. 1968 Jan 19; 159(3812):263-73.
- Scanning electrochemical microscopy study of ion annihilation electrogenerated chemiluminescence of rubrene and [Ru(bpy)3]2+.[J Am Chem Soc. 2012]Scanning electrochemical microscopy study of ion annihilation electrogenerated chemiluminescence of rubrene and [Ru(bpy)3]2+.Rodríguez-López J, Shen M, Nepomnyashchii AB, Bard AJ. J Am Chem Soc. 2012 Jun 6; 134(22):9240-50. Epub 2012 May 24.
- A simple and convenient electrochemiluminescence immunoassay by using gold nanoparticles as both label and co-reactant.[Bioelectrochemistry. 2020]A simple and convenient electrochemiluminescence immunoassay by using gold nanoparticles as both label and co-reactant.Zhao C, Niu L, Wang X, Sun W. Bioelectrochemistry. 2020 Oct; 135:107585. Epub 2020 Jun 20.
- Planning Implications Related to Sterilization-Sensitive Science Investigations Associated with Mars Sample Return (MSR).[Astrobiology. 2022]Planning Implications Related to Sterilization-Sensitive Science Investigations Associated with Mars Sample Return (MSR).Velbel MA, Cockell CS, Glavin DP, Marty B, Regberg AB, Smith AL, Tosca NJ, Wadhwa M, Kminek G, Meyer MA, et al. Astrobiology. 2022 Jun; 22(S1):S112-S164. Epub 2022 May 19.
- Electrochemiluminescence Method - StatPearlsElectrochemiluminescence Method - StatPearls
Your browsing activity is empty.
Activity recording is turned off.
See more...