158Flash Glucose Monitoring System for Diabetes

and .

Published online: June 1, 2017.

Summary

  • The FreeStyle Libre Pro is a glucose monitoring system intended to replace finger-stick tests; it uses a sensor implanted in the arm that a health care provider scans with a specialized reader for a record of glucose levels, trends, and patterns in people with diabetes.
  • The available evidence suggests that the performance and accuracy of the FreeStyle Libre Pro system is comparable to conventional methods of glucose monitoring, that it may reduce periods of hypoglycemia, or low blood sugar, and that it has high patient satisfaction.
  • Although the FreeStyle Libre Pro system is approved for use in Canada, it is not yet commercially available here.
  • Canadian pricing is not available, but cost-effectiveness analyses conducted in the UK and other European countries have concluded that it is cost-effective compared with the self-monitoring of blood glucose.
  • A consumer version of the FreeStyle Libre Pro system intended for personal use by patients with diabetes is currently awaiting regulatory approval in the US.
Image flash_glucose-diabetes-01-13f1

Image courtesy of Image courtesy of Abbott Diabetes Care

Issue

People living with diabetes must regularly monitor their glucose (blood sugar) levels to confirm and treat low blood sugar, or hypoglycemia, to adjust their dosage of insulin, and to provide information to their health care providers about their diabetes management.1 The frequency of self-monitoring blood glucose, which typically involves finger sticks and testing using capillary blood glucose monitors, should be individualized. It is recommended, though, that in people using insulin more than once a day, self-monitoring should be undertaken at least three times daily and before and after meals, especially if using rapid-acting, mealtime insulin.1 Some people with hard-to-control type 1 diabetes may opt for continuous glucose monitoring with devices that continuously measure glucose levels using a sensor implanted under the skin.1 Other people may opt for continuous monitoring because they find the number of finger sticks required for self-monitoring to be painful and inconvenient.

A new approach to glucose monitoring, called flash glucose monitoring, is becoming available and may offer people with type 1 and type 2 diabetes a less painful and more convenient option when intensive glucose monitoring is needed.

The Technology

The FreeStyle Libre Pro system (Abbott Diabetes Care, Alameda, California) is a flash glucose monitoring system that provides a record of an individual’s glucose levels, trends, and patterns for up to 14 days.2,3 The FreeStyle system generates an ambulatory glucose profile — a novel way of assessing glucose levels on a continuous 24-hour basis that shows how day-to-day decisions and behaviours impact the control of blood sugar levels.4

Similar to continuous glucose monitoring, flash glucose monitoring relies on the measurement of glucose concentrations in the interstitial fluid — the fluid surrounding cells — rather than directly measuring glucose levels in the blood.3 The FreeStyle Libre Pro system is intended for professional use only; i.e., the person with diabetes does not interact with the system. Rather, the sensor attachment is inserted and scanned by a health care provider to receive the data retroactively.3,5 A consumer version of the system, the FreeStyle Libre, is inserted and scanned by the person with diabetes, rather than the health care provider. It is currently under review by the US FDA2,6 and is already available in more than 35 countries worldwide (Jessica Sachariason, Abbott Diabetes Care, Alameda, CA: personal communication, 2017 Mar 10).

The FreeStyle Libre Pro system consists of a small, round, disposable, and water-resistant sensor approximately the size of two stacked quarters that is applied to the back of the individual’s upper arm by a health care provider in a clinic setting.2 The sensor continuously measures glucose in the interstitial fluid through a small, 5 mm-long-by-0.4 mm-wide filament inserted just under the skin. The sensor records glucose levels every 15 minutes, capturing up to 1,340 glucose results for up to 14 days, and displays current, eight-hour historic and trend glucose data when scanned.2 Information about the individual’s glucose levels is stored in the sensor which, in turn, is accessed by passing a specialized, hand-held reader over the sensor when the individual returns to the clinic or physician’s office.2

Flash glucose monitoring differs from continuous glucose monitoring in that information about the person’s glucose levels and trends is available from the FreeStyle Libre Pro system only when the sensor is scanned. In comparison, continuous glucose monitoring systems monitor and send information to a device or display monitor, without interruption, throughout the day, and can alert a user if glucose levels are outside a pre-set limit.1

An important distinction of the FreeStyle Libre Pro system compared with continuous glucose monitoring is that it is factory-calibrated and does not need daily calibration (or finger sticks) by the user. It also has a longer period of sensor use, 14 days versus 7 days, reads results in seconds, and can be used by people with type 2 diabetes.7,8 Two other key differences are that the FreeStyle Libre Pro system is not equipped with alarms (or notifications) that will sound in the event of out-of-range glucose values, and it lacks the ability to connect with insulin pump systems that continuously deliver short-acting insulin.7,9

In the past, continuous glucose monitoring systems have been used primarily by people with type 1 diabetes; however, complications arising from suboptimal glucose control affects both those with type 1 and type 2 diabetes.10 The FreeStyle Libre Pro system has the potential to address an unmet need for more intensive glucose monitoring in patients with type 1 or type 2 diabetes who require more vigilant glucose monitoring but who find the currently available continuous glucose monitoring systems too difficult, expensive, or cumbersome to use.

Availability

The FreeStyle Libre Pro system received a Class III medical device licence from Health Canada in November of 2016.11 The approval extends to three components of the system: the reader kit, sensor kit, and software.11 The FreeStyle Libre Pro system does not appear to be commercially available in Canada yet; however, health care professionals can order it on-line directly from the manufacturer’s US website (Jessica Sachariason: personal communication, 2017 Mar).3,7 The consumer version of the FreeStyle Libre Pro system is not currently licensed by Health Canada.

In September 2016, the FDA approved the FreeStyle Libre Pro system for people with diabetes aged 18 years and older.12,13 A consumer version of the FreeStyle Libre system intended for personal use is currently under review by the FDA.2,8,14 In Europe, the consumer version of the FreeStyle Libre system received a CE mark authorizing its use in adults in September of 2014, and a subsequent CE mark in February 2016 authorizing use in children and teens (aged four to 17 years).6,15

Cost

Canadian pricing information for the FreeStyle Libre Pro system is not available.

In the US, the FreeStyle Libre Pro reader costs US$65 and each individual sensor costs US$60.8 In the UK, it is reported that the cost for a starter pack of the FreeStyle Libre consumer system, which includes one reader and two sensors, is approximately £170.16

Who Might Benefit?

To date, continuous glucose monitoring has been used primarily by well-informed people with type 1 diabetes who have been trained on how to use it.7 In contrast, flash glucose monitoring may be a more accessible technology for people with type 2 diabetes, as it may enable appropriately trained people to monitor their glucose levels regularly after a meal or physical exercise, pending the availability of the consumer version of the system.7 It has been suggested that the FreeStyle Libre Pro system will be used primarily by people with type 2 diabetes; it could also benefit long-term users of continuous glucose monitors who find the alarms with most of these systems to be bothersome.7,17

The FreeStyle Libre Pro system may also be of benefit to people who are unable or unwilling to check their blood sugar using finger sticks because of pain and discomfort, as the system does not need routine finger-stick testing. This may be especially beneficial for children with type 1 diabetes; however, in North America, the system is not currently approved for use in people younger than 18 years of age.

Current Practice

Measurement of glycated hemoglobin (A1C) is the preferred standard for assessing long-term blood glucose control in people with both type 1 and type 2 diabetes.10,18 A major drawback of A1C is that it does not provide information about fluctuations in glucose levels; i.e., it is possible for A1C to remain within the target range while glucose levels fluctuate between periods of low (hypoglycemia) and high (hyperglycemia) levels throughout the day.18

“The FreeStyle Libre Pro system may also be of benefit to people who are unable or unwilling to check their blood sugar using finger sticks because of pain and discomfort…”

Self-monitoring of blood glucose using finger sticks and capillary blood glucose testing provides only discrete “snapshots” of glycemic control at the test times. Self-monitoring is further limited by its time-intensive nature and the discipline required to maintain records of daily blood glucose results over the long-term.18 The timing and frequency of self-monitoring of blood glucose depends on the type of diabetes and the treatment prescribed. It also depends on the individual’s need for information about glucose levels and the ability to use the test information to modify behaviour or adjust treatment.1

Continuous glucose monitoring has been used in people with type 1 diabetes to provide comprehensive, continuous glycemic profiles, which can improve glycemic control and reduce hypoglycemia.1,18 Notably though, continuous glucose monitoring systems still require self-monitoring for calibration of the system.1 Self-monitoring must be done at least every 12 hours and also in “real time” to confirm interstitial fluid measurements before making changes to therapy or treating suspected hypoglycemia.1

The Evidence

Nine published studies of the FreeStyle Libre system, used in adults, were identified.9,1926 Although it is the FreeStyle Libre Pro system that has been approved for use in Canada, it appears these studies evaluated the consumer version (the FreeStyle Libre system), which is designed for personal use and has been available in Europe since 2014.6 Four of the studies were conducted on people with type 1 diabetes,9,19,20,26 four on a mixed population of people with type 1 or type 2 diabetes,2124 and one study was on people undergoing cardiac surgery.25 Six of the studies were funded by the manufacturer9,19,2124 and three were investigator-initiated.20,25,26

Five of the nine studies evaluated the performance and accuracy of the FreeStyle Libre system compared with other methods of measuring glucose.20,21,2426 The other four studies reported clinical outcomes (for example hypoglycemia, A1C, patient satisfaction, and quality of life).9,19,22,23 Of these, three were observational, single-arm studies9,22,23 and one was a prospective, open-label, multi‐centre, randomized controlled trial (RCT), the IMPACT study.19 The primary outcome in the IMPACT study was the change in the amount of time spent in hypoglycemia from baseline to six months in 241 patients with well-controlled type 1 diabetes who were randomized to either the FreeStyle Libre system or to conventional self-monitoring of blood glucose.19

We also identified two published studies of the FreeStyle Libre system used in children with type 1 diabetes.27,28 One trial was funded by the manufacturer27 and the other was independently funded.28 The first study compared device performance and patient satisfaction between the FreeStyle Libre system and capillary blood glucose testing in the same children (age range: one to 14 years).27 The second was a pilot study to evaluate the feasibility and acceptability of the ambulatory glucose profile in children aged four to 17 years who were initiated on the FreeStyle Libre system.28

Performance and Accuracy

The performance and accuracy of the FreeStyle Libre system in adults and children compared with self-monitoring of blood glucose or capillary blood glucose testing, venous blood glucose analysis, continuous glucose monitoring, or intravascular microdialysis continuous glucose monitoring, are summarized in Table 1. All the studies in Table 1 reported the overall mean absolute relative difference (MARD), expressed as a percentage. A lower MARD value indicates smaller differences, or per cent error, between methods. The overall MARD between the FreeStyle Libre system compared with capillary blood glucose testing ranged from 10.0%24 to 13.9%;27 the MARD was 16.6% compared with self-monitored blood glucose,20 and 18.1% compared with continuous glucose monitoring.20 When directly compared with reference venous blood glucose results, the FreeStyle Libre system had a MARD of 13.2%, while the continuous glucose monitoring systems evaluated had MARDs of 16.8% (Dexcom G4 Platinum) and 21.4% (Medtronic MiniMed 640G).26 It should be noted that these MARD results were obtained under controlled conditions and that the accuracy and performance of these continuous glucose monitoring systems may vary considerably during periods of rapidly changing glucose levels, such as during exercise or after meals.26,29

Table 1

Table 1

Performance and Test Characteristics of the FreeStyle Libre Flash Glucose Monitoring System

“The average amount of time participants experienced hypoglycemia was statistically significantly lower at six months for people who used the FreeStyle Libre system (1.24 fewer hours per day) compared with those who self-monitored their blood glucose.”

The results of error grid analyses were also reported. This method describes the clinical accuracy of a test method compared with a reference method based on paired glucose samples and the clinical significance of the differences between them.30,31 Researchers found that between 62.4% and 94.0% of the paired samples of the FreeStyle Libre system test method and the reference test method were within zone A — meaning that test values deviated less than 20% from the reference value and have no clinical implications. Between 91.6% and 100% of paired samples were within zones A and B — meaning that, although the difference may be more than 20%, no treatment is indicated, or treatment would have little or no effect on clinical outcome).30,31

There was also a high level of agreement between the FreeStyle Libre system and the other test methods, ranging from 76% agreement with continuous glucose monitoring to 95% agreement with venous blood glucose. Two studies reported the mean lag time in glucose levels associated with the FreeStyle Libre system compared with that of venous blood glucose, which ranged from 3.1 minutes to 4.5 minutes.21,24 In the study of patients undergoing cardiac surgery, it was reported that, although both systems were reliable and no complications occurred, the intravascular microdialysis continuous glucose monitoring system was more accurate than the FreeStyle Libre system (which repeatedly measured a lower glucose value).25

Hypoglycemia

The primary outcome in the IMPACT study was the change in the amount of time in hypoglycemia — defined as hours per day, with a glucose level of less than 3.9 mmol/L (70 mg/dL) from baseline to six months.19 The average amount of time participants experienced hypoglycemia was statistically significantly lower at six months for people who used the FreeStyle Libre system (1.24 fewer hours per day) compared with those who self-monitored their blood glucose.19 Analyses of day and night results found that the amount of time spent below all hypoglycemic thresholds and the number of hypoglycemic episodes were also significantly reduced in patients who used the FreeStyle Libre system compared with those who self-monitored their blood glucose.19

A single-arm study that prospectively assessed the impact of introducing the FreeStyle Libre system to people with type 1 diabetes over a 16-week period reported that hypoglycemic episodes, recorded and self-reported, were reduced: from 17 episodes in the first two weeks of the study to 12 episodes in the final two weeks of the study.9 Another single-arm study of people with type 1 or type 2 diabetes that assessed aggregate data from ambulatory glucose profiles obtained using the FreeStyle Libre system found that, despite having similar A1C values, patients with type 2 diabetes had different glycemic characteristics and a different risk of hypoglycemia than those with type 1 diabetes.22 It was noted that these data, which have important implications for therapeutic adjustments, were obtained within a short time and with minimal provider and patient input using the FreeStyle Libre system.22

In one of the studies of the FreeStyle Libre system in children with type 1 diabetes, sensor data were used to retrospectively determine glycemic variability in the children during the course of the study.27 Overall, the children were in the target glucose range of 3.9 mmol/L to 10.0 mmol/L (70 mg/dL to 180 mg/dL) approximately 50% of the time, or an average of 12.1 hours per day. This range was the same both during the day and at night, regardless of whether insulin was administered by insulin pump or by multiple daily injections.27

Glycated Hemoglobin

There are conflicting reports about the effects of the FreeStyle Libre system on A1C. In the IMPACT study, although time spent in hypoglycemia was significantly reduced over the six-month period, A1C values remained essentially unchanged and were not statistically significantly different in participants who used the FreeStyle Libre system compared with those who self-monitored their blood glucose.19 In contrast, two single-arm studies without control groups reported statistically significant reductions in A1C values following the introduction of the FreeStyle Libre system.9,23

Patient Satisfaction

In the adult studies that reported patient satisfaction, all found a high level of patient satisfaction with the FreeStyle Libre system.9,19,21,23,24 In the IMPACT study, both patient satisfaction and overall treatment satisfaction were statistically significantly higher with the FreeStyle Libre system compared with self-monitoring of blood glucose, although no difference was detected between groups when they were asked about diabetes distress, fear of hypoglycemia, or worry scores.19

In the study comparing the FreeStyle Libre system and capillary blood glucose monitoring in children, sensor application, wear and use of the device, and comparison with the self-monitoring of blood glucose were rated favourably by most children and caregivers (84.3% to 100%).27 In the second study in children, the FreeStyle Libre sensor was maintained in place for the full 14-day period in 65% of the children.28 With the exception of minor discomfort, the system was well-accepted by the majority of children and parents, and would be used again by 60% of the study population.28

Patient Factors

Two studies, one in adults21 and the other in children,27 reported that factors such as age, body weight or body mass index, sex, A1C level, method of insulin administration, and time of use (day versus night) did not affect sensor accuracy.

A subgroup analysis of the IMPACT study reported in abstract form evaluated whether younger adults (less than 25 years of age) were less likely than older adults (25 years of age and older) to regularly use glucose monitoring devices.32 After six months of use, there were no significant differences between these subgroups of participants in the number of scans, amount of sensor wear, or sensor-derived glucose results.32 In addition, there was no significant interaction between participant age and time in hypoglycemia, in keeping with the results for the overall study population.32

The IMPACT study also reported on the average number of daily finger-stick tests performed by all participants.19 In users of the FreeStyle Libre system, testing was reduced from an average of 5.5 finger-stick tests per day during the baseline period to 0.5 tests per day (one test every two to five days) during the treatment phase, compared with 5.6 tests per day with self-monitoring of blood glucose.19 The mean number of sensor scans per day in users of the FreeStyle Libre system was 15.1 during the treatment phase.19 A recent report based on real-world use of the FreeStyle Libre system found that people using the system checked their glucose levels an average of 16 times per day and that average glucose levels, A1C levels, the time spent in hypoglycemia, and instances of hyperglycemia all decreased as scan rates increased.33

Quality of Life

Various patient-reported outcomes (using instruments such as the Hypoglycemia Fear Survey, Diabetes Treatment and Satisfaction Questionnaire, Diabetes Distress Scale [DDS], and the Diabetes Quality of Life questionnaire) were reported in the IMPACT study.19 The results were inconclusive because some differences between measures were statistically significant, whereas others were not.19

In one single-arm study, quality of life was measured using the DDS; reductions in the mean scale score, and emotional burden and regimen-related distress sub‐scores, were reported 16 weeks after introducing the FreeStyle Libre system.9

Cost-Effectiveness

Three cost-effectiveness analyses, all reported in abstract form, were identified that compared the FreeStyle Libre system with conventional self-monitoring of blood glucose, or continuous blood glucose testing, in the UK and other European countries.3436 Overall, compared with self-monitoring or continuous blood glucose testing, the FreeStyle Libre system was cost-effective.

Safety

None of the studies reported serious device-related safety issues or malfunctions associated with the FreeStyle Libre system. In general, most of the adverse events reported for the FreeStyle Libre system were associated with the sensor and/or adhesive (e.g., local irritation, allergy events, itching, redness, rash, and insertion-site symptoms).19,21,22,27

“…caution should be taken when the system is used with other implanted medical devices such as pacemakers, and some individuals may be sensitive to the adhesive used on the sensor.”

The clinical study that formed the basis for the FDA approval of the FreeStyle Libre Pro system37 found that, when the device indicated glucose values at or below 60 mg/dL (3.3 mmol/L), 40% of the time those values were actually in the range of 81 mg/dL to 160 mg/dL (4.5 mmol/L to 8.9 mmol/L).5 Interpreting readings should therefore be based only on trends and patterns analyzed over time, using the reports available (per the intended use).5

Contraindications to the use of the FreeStyle Libre Pro system are that the sensor must be removed before having an MRI, CT scan, or high-frequency electrical heat (diathermy) treatment, as the effects of these interventions on the performance of the FreeStyle Libre Pro system have not been evaluated.5 Furthermore, caution should be taken when the system is used with other implanted medical devices such as pacemakers, and some individuals may be sensitive to the adhesive used on the sensor.5

Concurrent Developments

Various non-invasive glucose monitors are currently in development or are awaiting regulatory approval. The GlucoTrack DF-F (Integrity Applications, Ashkelon, Israel) works in a manner similar to that of a pulse oximeter. It consists of a small sensor that is clipped to the earlobe and connected to a hand-held control and display unit.38 The NovioSense (Novio Tech Campus, Nijmegen, Netherlands) is a small glucose sensor placed in the eye to continuously measure glucose levels in the tears.38 It is anticipated that it may enter the market in 2018. The OrSense NBM-200G (OrSense, Petach Tikva, Israel) is a finger sensor that uses changes in the red and near-infrared spectrum of light passed through the finger to correlate with, and detect changes in, glucose concentrations.38

Implementation Issues

The FreeStyle Libre Pro system generates a large volume of data that may be overwhelming, although presenting the data as an ambulatory glucose profile may mitigate this issue for some users.7,39

Wearing the sensor could cause skin irritations or it could be perceived as unsightly.16 Patient satisfaction and sensor wear and tear are important considerations, as one study reported that continuous glucose monitoring sensors were only worn 57% of the time by adults with type 1 diabetes enrolled in an RCT to assess impaired awareness of hypoglycemia.40 One of the single-arm studies of the FreeStyle Libre system reported that the sensor did not record any results in five patients (6.8%) despite being in place for 14 days.22 In addition, 12 sensors (16.4%) had less than five days of data — either because they fell off or were removed early.22

Because the FreeStyle Libre Pro sensor must be replaced every 14 days by a health care provider, this may impact the number of clinician office visits for the person using the system.

For both continuous and flash glucose monitoring systems, because measurements are taken from interstitial fluid and not blood, there is a lag time in measurement of up to 15 minutes.17,41 The lag time is longest if the glucose levels are changing rapidly (e.g., after eating, dosing of insulin, or exercising); therefore, it is still necessary to do a finger-stick and capillary blood glucose test if a change in treatment is being contemplated.

There are no reusable components in the FreeStyle Libre Pro system, other than the reader. The sensor should not be reused and is not suitable for re-sterilization.5 The sensor reader should be cleaned between patient use.5

As there is no patient interaction with the system, patients do not require training on the calibration or use of the system. This would change if the consumer version of the FreeStyle Libre system becomes available in Canada.

With the FreeStyle Libre Pro system, the clinic or physician’s office is required to purchase only one FreeStyle Libre Pro reader, which can be used for multiple patients.

Final Remarks

Flash glucose monitoring has been described as occupying a clinical space between continuous glucose monitoring and the self-monitoring of blood glucose.18 In principle, flash glucose monitoring using the FreeStyle Libre Pro system could replace self-monitoring with finger sticks and capillary blood glucose testing, as it offers a more convenient and pain-free alternative, especially if frequent testing is required. It is important to note that although the need for routine finger sticks is diminished, they are still required if a change in therapy (such as an adjustment of insulin dose) is being contemplated or if hypoglycemia is suspected.

Because of the pain and discomfort of finger sticks and the need to wake up in the night to test blood glucose levels when nighttime hypoglycemia is a concern, it is especially difficult to have frequent self-monitoring in children.15 The FreeStyle Libre Pro system may be beneficial for children with type 1 diabetes who require intensive glucose monitoring; however, in North America, it is not currently approved for use in children.

Box Icon

Box

Methods — Literature Search Strategy.

Acknowledgments

CADTH thanks the external reviewers who kindly provided comments on an earlier draft of this bulletin.

Footnotes

Disclaimer: The information in this document is intended to help Canadian health care decision-makers, health care professionals, health systems leaders, and policy-makers make well-informed decisions and thereby improve the quality of health care services. While patients and others may access this document, the document is made available for informational purposes only and no representations or warranties are made with respect to its fitness for any particular purpose. The information in this document should not be used as a substitute for professional medical advice or as a substitute for the application of clinical judgment in respect of the care of a particular patient or other professional judgment in any decision-making process. The Canadian Agency for Drugs and Technologies in Health (CADTH) does not endorse any information, drugs, therapies, treatments, products, processes, or services.

While CADTH has taken care to ensure that the information prepared by it in this document is accurate, complete, and up-to-date as at the applicable date the material was first published by CADTH, CADTH does not make any guarantees to that effect. CADTH does not guarantee and is not responsible for the quality, currency, propriety, accuracy, or reasonableness of any statements, information, or conclusions contained in any third-party materials used in preparing this document. The views and opinions of third parties published in this document do not necessarily state or reflect those of CADTH.

CADTH is not responsible for any errors, omissions, injury, loss, or damage arising from or relating to the use (or misuse) of any information, statements, or conclusions contained in or implied by the contents of this document or any of the source materials.

This document may contain links to third-party websites. CADTH does not have control over the content of such sites. Use of third-party sites is governed by the third-party website owners’ own terms and conditions set out for such sites. CADTH does not make any guarantee with respect to any information contained on such third-party sites and CADTH is not responsible for any injury, loss, or damage suffered as a result of using such third-party sites. CADTH has no responsibility for the collection, use, and disclosure of personal information by third-party sites.

Subject to the aforementioned limitations, the views expressed herein are those of CADTH and do not necessarily represent the views of Canada’s federal, provincial, or territorial governments.

This document is prepared and intended for use in the context of the Canadian health care system. The use of this document outside of Canada is done so at the user’s own risk.

This disclaimer and any questions or matters of any nature arising from or relating to the content or use (or misuse) of this document will be governed by and interpreted in accordance with the laws of the Province of Ontario and the laws of Canada applicable therein, and all proceedings shall be subject to the exclusive jurisdiction of the courts of the Province of Ontario, Canada.

The copyright and other intellectual property rights in this document are owned by CADTH and its licensors. These rights are protected by the Canadian Copyright Act and other national and international laws and agreements. You are permitted to make copies of this document for non-commercial purposes only, provided it is not modified when reproduced and appropriate credit is given to CADTH and its licensors.

About CADTH: CADTH is an independent, not-for-profit organization responsible for providing Canada’s health care decision-makers with objective evidence to help make informed decisions about the optimal use of drugs, medical devices, diagnostics, and procedures in our health care system.

Funding: CADTH receives funding from Canada’s federal, provincial, and territorial governments, with the exception of Quebec.

References

1.
Canadian Diabetes Association Clinical Practice Guidelines Expert Committee. Berard LD, Blumer I, Houlden R, Miller D, Woo V. Monitoring glycemic control. Can J Diabetes. 2013 Apr;37(Suppl 1):S35–S39. [PubMed: 24070960]
2.
Abbott receives FDA approval for the FreeStyle Libre Pro™ system, a revolutionary diabetes sensing technology for healthcare professionals to use with their patients [Internet]. Abbott Park (IL): Abbott; Sep 28, 2016. [cited 2017 Feb 15]. Available from: http://abbott​.mediaroom​.com/2016-09-28-Abbott-Receives-FDA-Approval-for-the-FreeStyle-Libre-Pro-System-a-Revolutionary-Diabetes-Sensing-Technology-for-Healthcare-Professionals-to-Use-with-Their-Patients.
3.
FreeStyle Libre Pro [Internet]. Alameda (CA): Abbott Diabetes Care; 2016. [cited 2017 Feb 15]. Available from: http://www​.freestylelibrepro.us/
4.
Kalra S, Gupta Y. Ambulatory glucose profile: Flash glucose monitoring. J Pak Med Assoc [Internet]. Dec, 2015. pp. 1360–2. [cited 2017 Jan 30] Available from: http://jpma​.org.pk/full_article_text​.php?article_id=7566. [PubMed: 26627526]
5.
FreeStyle Libre Pro flash glucose monitoring system: operator’s manual [Internet]. Alameda (CA): Abbott Diabetes Care Inc.; Sep 23, 2016. [cited 2017 Feb 15]. Available from: http://www​.accessdata​.fda.gov/cdrh_docs/pdf15/P150021C.pdf.
6.
Tucker ME. Medscape [Internet]. New York (NY): WebMD LLC; Sep 5, 2014. No more finger pricks? EU okays new glucose monitoring system. [cited 2017 Feb 15]. Available from: http://www​.medscape.com​/viewarticle/831203 Free subscription required.
7.
Heinemann L, Freckmann G. CGM Versus FGM; Or, continuous glucose monitoring is not flash glucose monitoring. J Diabetes Sci Technol [Internet]. Sep, 2015. pp. 947–50. [cited 17 A.D. Jan 30] Available from: https://www​.ncbi.nlm​.nih.gov/pmc/articles/PMC4667350/ [PMC free article: PMC4667350] [PubMed: 26330484]
8.
Tucker ME. Medscape [Internet]. New York (NY): WebMD LLC; Sep 28, 2016. FDA approves Abbotts FreeStyle Libre Pro system for diabetes. [cited 2017 Feb 15]. Available from: http://www​.medscape.com​/viewarticle/869386 Free subscription required.
9.
Dover AR, Stimson RH, Zammitt NN, Gibb FW. Flash Glucose Monitoring Improves Outcomes in a Type 1 Diabetes Clinic. J Diabetes Sci Technol. 2016 Jul 26; [PMC free article: PMC5478014] [PubMed: 27464754]
10.
Canadian Diabetes Association Clinical Practice Guidelines Expert Committee. Imran SA, Rabasa-Lhoret R, Ross S. Targets for glycemic control. Can J Diabetes. 2013 Apr;37(Suppl 1):S31–S34. [PubMed: 24070959]
11.
Medical Devices Active Licence Listing (MDALL) [Internet]. Ottawa: Health Canada; 2016. FreeStyle Libre Pro flash glucose monitoring system. [cited 2017 Feb 10]. Available from: https:​//health-products​.canada.ca/mdall-limh/index-eng.jsp.
12.
Premarket approval (PMA) [Internet]. Silver Spring (MD): U.S. Food and Drug Administration; Mar 27, 2017. FreeStyle Libre Pro flash glucose monitoring system: supplement record. [cited 2017 Feb 15]. Available from: https://www​.accessdata​.fda.gov/scripts/cdrh​/cfdocs/cfpma/pma.cfm?id=P150021S003.
13.
Premarket approval (PMA) [Internet]. Silver Spring (MD): U.S. Food and Drug Administration; Mar 27, 2017. FreeStyle Libre Pro flash glucose monitoring system. [cited 2017 Feb 15]. Available from: https://www​.accessdata​.fda.gov/scripts/cdrh​/cfdocs/cfpma/pma.cfm?id=P150021.
14.
Abbott announces positive results from groundbreaking study of FreeStyle Libre system for people with type 1 diabetes [Internet]. New Orleans: PRNewswire; Jun 13, 2016. [cited 2017 Feb 15]. Available from: http://www​.prnewswire​.com/news-releases/abbott-announces-positive-results-from-groundbreaking-study-of-freestyle-libre-system-for-people-with-type-1-diabetes-300283469.html.
15.
Children in Europe now have access to first-of-its-kind diabetes sensing technology - eliminates routine finger pricks [Internet]. Abbott Park (IL): PR Newswire; Feb 4, 2017. [cited 2017 Feb 21]. Available from: http://www​.prnewswire​.com/news-releases/children-in-europe-now-have-access-to-first-of-its-kind-diabetes-sensing-technology---eliminates-routine-finger-pricksi-300214856.html.
16.
Flash glucose monitoring [Internet]. London: Diabetes UK; 2017. [cited 2017 Feb 15]. Available from: https://www​.diabetes​.org.uk/Guide-to-diabetes​/Managing-your-diabetes​/Testing/Flash-glucose-monitoring/
17.
Rodbard D. Continuous glucose monitoring: A review of successes, challenges, and opportunities. Diabetes Technol Ther [Internet]. Feb, 2016. pp. S2-3–S2-13. [cited 2017 Jan 30] Available from: https://www​.ncbi.nlm​.nih.gov/pmc/articles/PMC4717493. [PMC free article: PMC4717493] [PubMed: 26784127]
18.
Evans M. Current methods of assessing blood glucose control in diabetes. British Journal of Diabetes. 2016;16(Suppl 1):S7–S9.
19.
Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kroger J, Weitgasser R. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet. 2016 Nov 5;388(10057):2254–63. [PubMed: 27634581]
20.
Bonora B, Maran A, Ciciliot S, Avogaro A, Fadini GP. Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes. J Endocrinol Invest. 2016 Dec;39(12):1391–9. [PubMed: 27287421]
21.
Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The performance and usability of a factory-calibrated flash glucose monitoring system. Diabetes Technol Ther [Internet]. Nov, 2015. pp. 787–94. [cited 2017 Jan 30] Available from: https://www​.ncbi.nlm​.nih.gov/pmc/articles/PMC4649725. [PMC free article: PMC4649725] [PubMed: 26171659]
22.
Distiller LA, Cranston I, Mazze R. First Clinical Experience with Retrospective Flash Glucose Monitoring (FGM) Analysis in South Africa: Characterizing Glycemic Control with Ambulatory Glucose Profile. J Diabetes Sci Technol. 2016 Nov;10(6):1294–302. [PMC free article: PMC5094324] [PubMed: 27154973]
23.
Ish-Shalom M, Wainstein J, Raz I, Mosenzon O. Improvement in Glucose Control in Difficult-to-Control Patients With Diabetes Using a Novel Flash Glucose Monitoring Device. J Diabetes Sci Technol. 2016 Nov;10(6):1412–3. [PMC free article: PMC5094332] [PubMed: 27277660]
24.
Ji L, Guo X, Guo L, Ren Q, Yu N, Zhang J. A Multicenter Evaluation of the Performance and Usability of a Novel Glucose Monitoring System in Chinese Adults With Diabetes. J Diabetes Sci Technol. 2017 Mar;11(2):290–5. [PMC free article: PMC5478018] [PubMed: 27559031]
25.
Schierenbeck F, Franco-Cereceda A, Liska J. Accuracy of 2 Different Continuous Glucose Monitoring Systems in Patients Undergoing Cardiac Surgery: Intravascular Microdialysis Versus Subcutaneous Tissue Monitoring. J Diabetes Sci Technol. 2016 Jun 1;11(1):108–16. [PMC free article: PMC5375060] [PubMed: 27257168]
26.
Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann P, Elsayed H, et al. Evaluation of Subcutaneous Glucose Monitoring Systems under Routine Environmental Conditions in Patients with Type 1 Diabetes. Diabetes Obes Metab. 2017 Feb 16; Epub ahead of print. [PubMed: 28205324]
27.
Edge J, Acerini C, Campbell F, Hamilton-Shield J, Moudiotis C, Rahman S, et al. An alternative sensor-based method for glucose monitoring in children and young people with diabetes. Arch Dis Child [Internet]. Jan 30, 2017. [cited 2017 Feb 13]. Available from: http://adc​.bmj.com/content​/archdischild/early​/2017/01/30/archdischild-2016-311530.full.pdf. [PMC free article: PMC5466923] [PubMed: 28137708]
28.
Rai S, Hulse A, Kumar P. Feasibility and acceptability of ambulatory glucose profile in children with Type 1 diabetes mellitus: A pilot study. Indian J Endocrinol Metab. 2016 Nov;20(6):790–4. [PMC free article: PMC5105562] [PubMed: 27867881]
29.
Taleb N, Emami A, Suppere C, Messier V, Legault L, Chiasson JL, et al. Comparison of two continuous glucose monitoring systems, Dexcom G4 Platinum and Medtronic Paradigm Veo Enlite System, at rest and during exercise. Diabetes Technol Ther. 2016 Sep;18(9):561–7. [PubMed: 27356172]
30.
Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care [Internet]. Aug, 2000. pp. 1143–8. [cited 2017 Feb 16] Available from: http://care​.diabetesjournals​.org/content/diacare/23/8/1143​.full.pdf. [PubMed: 10937512]
31.
Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL. Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data. Diabetes Care [Internet]. Aug, 2004. pp. 1922–8. [cited 2017 Feb 15] Available from: http://care​.diabetesjournals​.org/content/diacare/27/8/1922​.full.pdf. [PubMed: 15277418]
32.
Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Matthaei S, Weitgasser R. Using novel flash glucose-sensing technology for 6 months results in a high rate of concordance by young adults with type 1 diabetes [abstract]. Diabetologia; Presented at 52nd Annual Meeting of the European Association for the Study of Diabetes, EASD; 2016 Sep 12–16; Munich, Germany. 2016. p. S420.
33.
Ford O. Freestyle Libre system gains ground over traditional glucose monitoring methods. Medical Device Daily. 2017 Feb 21;21(34):1, 4.
34.
Li H, Bilir SP, Wehler EA, Hellmund R, Munakata J. Cost effectiveness analysis of a flash glucose monitoring system for type 2 diabetes (T2DM) patients receiving intensive insulin treatment in Europe [abstract]. Value Health; Presented at ISPOR 19th Annual European Congress; 2016 Oct 29–Nov 2; Vienna, Austria. 2016. p. A698.
35.
Li H, Bilir SP, Donga P, Samiian A, Munakata J. Cost Effectiveness Analysis of Flash Glucose Monitoring for Type 2 Diabetes Patients Receiving Insulin Treatment In The UK [abstract] Value Health. 2014 Nov;17(7):A351. [PubMed: 27200680]
36.
Bilir SP, Li H, Wehler EA, Hellmund R, Munakata J. Cost effectiveness analysis of a flash glucose monitoring system for type 1 diabetes (T1DM) patients receiving intensive insulin treatment in Europe and Australia [abstract]. Value Health; Presented at ISPOR 19th Annual European Congress; 2016 Oct 29–Nov 2; Vienna, Austria. 2016. pp. A697–A698.
37.
Bailey T, Bode BW, Christiansen MP, Klaff LJ, Alva S. The performance and usability of a factory-calibrated flash glucose monitoring system. Diabetes Technol Ther. 2016;18:S18–S19. [PubMed: 27096931]
38.
New and emerging health technologies for diabetes [Internet]. Cologne (DEU): EuroScan International Network; Mar 17, 2016. [cited 2017 Feb 15]. Available from: https://www​.euroscan​.org/technologies/new-and-emerging-health-technologies-for-diabetes/
39.
Hammond P. Interpreting the ambulatory glucose profile. Br J Diab Vasc Dis. 2016;16:S10–S15.
40.
Little SA, Leelarathna L, Walkinshaw E, Tan HK, Chapple O, Lubina-Solomon A, et al. Recovery of hypoglycemia awareness in long-standing type 1 diabetes: a multicenter 2 × 2 factorial randomized controlled trial comparing insulin pump with multiple daily injections and continuous with conventional glucose self-monitoring (HypoCOMPaSS). Diabetes Care [Internet]. Aug, 2014. pp. 2114–22. [cited 2017 Mar 14] Available from: http://care​.diabetesjournals​.org/content/diacare/37/8/2114​.full.pdf. [PubMed: 24854041]
41.
Freckmann G, Pleus S, Schluter S, Heinemann L. Comment on “The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System” by Bailey et al. Diabetes Technol Ther. 2016 May;18(5):334–5. [PubMed: 26900898]
Copyright © CADTH 2017. You are permitted to reproduce this document for non-commercial purposes, provided it is not modified when reproduced and appropriate credit is given to CADTH.

Except where otherwise noted, this work is distributed under the terms of a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International licence (CC BY-NC-ND), a copy of which is available at https://creativecommons.org/licenses/by-nc-nd/4.0/

Bookshelf ID: NBK476439PMID: 29369576

Views

Related information

  • PMC
    PubMed Central citations
  • PubMed
    Links to PubMed

Similar articles in PubMed

See reviews...See all...

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...