Objective: The aim of the trial was to compare the time parameters for intubation with the use of the Macintosh (MAC) laryngoscope and Pentax AWS-S100 videolaryngoscope (AWS; Pentax Corporation, Tokyo, Japan) with and without chest compression (CC) by paramedics during simulated cardiopulmonary resuscitation in a cadaver model.
Methods: This was a randomized crossover cadaver trial. Thirty-five paramedics with no experience in videolaryngoscopy participated in the study. They performed intubation in two emergency scenarios: scenario A, normal airway without CC; scenario B, normal airway with continuous CC.
Results: The median time to first ventilation with the use of the AWS and the MAC was similar in scenario A: 25 (IQR, 22-27) seconds vs. 24 (IQR, 22.5-26) seconds (P=.072). A statistically significant difference in TTFV between AWS and MAC was noticed in scenario B (P=.011). In scenario A, the first endotracheal intubation (ETI) attempt success rate was achieved in 97.1% with AWS compared with 94.3% with MAC (P=.43). In scenario B, the success rate after the first ETI attempt with the use of the different intubation methods varied and amounted to 88.6% vs. 77.1% for AWS and MAC, respectively (P=.002).
Conclusions: The Pentax AWS offered a superior glottic view as compared with the MAC laryngoscope, which was associated with a higher intubation rate and a shorter intubation time during an uninterrupted CC scenario. However, in the scenario without CC, the results for AWS and MAC were comparable.
Copyright © 2016 Elsevier Inc. All rights reserved.