Towards Bacteria Counting in DI Water of Several Microliters or Growing Suspension Using Impedance Biochips

Biosensors (Basel). 2020 Jul 23;10(8):82. doi: 10.3390/bios10080082.

Abstract

We counted bacterial cells of E. coli strain K12 in several-microliter DI water or in several-microliter PBS in the low optical density (OD) range (OD = 0.05-1.08) in contact with the surface of Si-based impedance biochips with ring electrodes by impedance measurements. The multiparameter fit of the impedance data allowed calibration of the impedance data with the concentration cb of the E. coli cells in the range of cb = 0.06 to 1.26 × 109 cells/mL. The results showed that for E. coli in DI water and in PBS, the modelled impedance parameters depend linearly on the concentration of cells in the range of cb = 0.06 to 1.26 × 109 cells/mL, whereas the OD, which was independently measured with a spectrophotometer, was only linearly dependent on the concentration of the E. coli cells in the range of cb = 0.06 to 0.50 × 109 cells/mL.

Keywords: Escherichia coli; bacterial cell counting; biochips; electrical equivalent circuit model; impedance spectroscopy.

MeSH terms

  • Bacteria*
  • Bacterial Load / methods
  • Biosensing Techniques / methods*
  • Electric Impedance
  • Microelectrodes
  • Water Microbiology*