Disposable Non-Enzymatic Glucose Sensors Using Screen-Printed Nickel/Carbon Composites on Indium Tin Oxide Electrodes

Sensors (Basel). 2015 Dec 10;15(12):31083-91. doi: 10.3390/s151229846.

Abstract

Disposable screen-printed nickel/carbon composites on indium tin oxide (ITO) electrodes (DSPNCE) were developed for the detection of glucose without enzymes. The DSPNCE were prepared by screen-printing the ITO substrate with a 50 wt% nickel/carbon composite, followed by curing at 400 °C for 30 min. The redox couple of Ni(OH)₂/NiOOH was deposited on the surface of the electrodes via cyclic voltammetry (CV), scanning from 0-1.5 V for 30 cycles in 0.1 M NaOH solution. The DSPNCE were characterized by field-emission scanning electron microscopy (FE-SEM), X-ray photoelectron spectroscopy (XPS), and electrochemical methods. The resulting electrical currents, measured by CV and chronoamperometry at 0.65 V vs. Ag/AgCl, showed a good linear response with glucose concentrations from 1.0-10 mM. Also, the prepared electrodes showed no interference from common physiologic interferents such as uric acid (UA) or ascorbic acid (AA). Therefore, this approach allowed the development of a simple, disposable glucose biosensor.

Keywords: biosensor; disposable glucose sensor; nickel/carbon composite; non-enzymatic glucose sensor.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Biosensing Techniques / methods
  • Carbon / chemistry
  • Electrochemical Techniques / instrumentation*
  • Electrochemical Techniques / methods
  • Electrodes
  • Glucose / analysis*
  • Nickel / chemistry*
  • Surface Properties
  • Tin Compounds / chemistry*

Substances

  • Tin Compounds
  • indium tin oxide
  • Carbon
  • Nickel
  • Glucose