Warning: The NCBI web site requires JavaScript to function. more...
Department of Anesthesia, University of Iowa, College of Medicine, Iowa City 52242.
To evaluate the reliability of capnography in identifying esophageal intubation in the presence of a carbonated beverage in the stomach, we first investigated the amount of CO2 released from different carbonated beverages and antacids in a simulated stomach; next we measured the end-expired CO2 level during esophageal ventilation with a carbonated beverage in the stomachs of six swine. CO2 levels of approximately 20% were consistently observed in all carbonated beverages. The CO2 levels obtained with sodium bicarbonate, Maalox, and sodium citrate were 19.3%, 2.0%, and 0%, respectively. CO2 waveforms were observed during esophageal ventilation in five of six animals after intragastric administration of a carbonated beverage. An end-expired CO2 level of 2.5% or more was observed in two swine. The highest end-expired CO2 level measured was 5.3%. We conclude that although capnography is convenient and effective, it lacks all the attributes of an ideal monitor for detecting esophageal intubation.