Format

Send to:

Choose Destination
See comment in PubMed Commons below
Magn Reson Med. 2006 Aug;56(2):386-94.

High magnetic field water and metabolite proton T1 and T2 relaxation in rat brain in vivo.

Author information

  • 1Magnetic Resonance Research Center, Yale University, School of Medicine, New Haven, Connecticut 06520-8043, USA. robin.degraaf@yale.edu

Abstract

Comprehensive and quantitative measurements of T1 and T2 relaxation times of water, metabolites, and macromolecules in rat brain under similar experimental conditions at three high magnetic field strengths (4.0 T, 9.4 T, and 11.7 T) are presented. Water relaxation showed a highly significant increase (T1) and decrease (T2) with increasing field strength for all nine analyzed brain structures. Similar but less pronounced effects were observed for all metabolites. Macromolecules displayed field-independent T2 relaxation and a strong increase of T1 with field strength. Among other features, these data show that while spectral resolution continues to increase with field strength, the absolute signal-to-noise ratio (SNR) in T1/T2-based anatomical MRI quickly levels off beyond approximately 7 T and may actually decrease at higher magnetic fields.

Copyright 2006 Wiley-Liss, Inc.

PMID:
16767752
[PubMed - indexed for MEDLINE]