A prospective, randomized evaluation of a novel everolimus-eluting coronary stent: the PLATINUM (a Prospective, Randomized, Multicenter Trial to Assess an Everolimus-Eluting Coronary Stent System [PROMUS Element] for the Treatment of Up to Two de Novo Coronary Artery Lesions) trial

J Am Coll Cardiol. 2011 Apr 19;57(16):1700-8. doi: 10.1016/j.jacc.2011.02.016. Epub 2011 Apr 4.

Abstract

Objectives: We sought to evaluate the clinical outcomes with a novel platinum chromium everolimus-eluting stent (PtCr-EES) compared with a predicate cobalt chromium everolimus-eluting stent (CoCr-EES) in patients undergoing percutaneous coronary intervention (PCI).

Background: Randomized trials have demonstrated an excellent safety and efficacy profile for the CoCr-EES. The PtCr-EES uses the identical antiproliferative agent and polymer but with a novel platinum chromium scaffold designed for enhanced deliverability, vessel conformability, side-branch access, radiopacity, radial strength, and fracture resistance.

Methods: A total of 1,530 patients undergoing PCI of 1 or 2 de novo native lesions were randomized at 132 worldwide sites to CoCr-EES (n = 762) or PtCr-EES (n = 768). The primary endpoint was the 12-month rate of target lesion failure (TLF), the composite of target vessel-related cardiac death, target vessel-related myocardial infarction (MI), or ischemia-driven target lesion revascularization (TLR) in the per-protocol population (patients who received ≥1 assigned study stent), powered for noninferiority.

Results: The 12-month rate of TLF in the per-protocol population occurred in 2.9% versus 3.4% of patients assigned to CoCr-EES versus PtCr-EES, respectively (difference: 0.5%, 95% confidence interval: -1.3% to 2.3%, p(noninferiority) = 0.001, p(superiority) = 0.60). By intention-to-treat, there were no significant differences between CoCr-EES and PtCr-EES in the 12-month rates of TLF (3.2% vs. 3.5%, p = 0.72), cardiac death or MI (2.5% vs. 2.0%, p = 0.56), TLR (1.9% vs. 1.9%, p = 0.96), or Academic Research Consortium definite or probable stent thrombosis (0.4% vs. 0.4%, p = 1.00).

Conclusions: In this large-scale, prospective, single-blind randomized trial, a novel PtCr-EES was noninferior to the predicate CoCr-EES for TLF, with nonsignificant differences in measures of safety and efficacy through 12-month follow-up after PCI.

Trial registration: ClinicalTrials.gov NCT00823212.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Coronary Artery Disease / drug therapy*
  • Coronary Artery Disease / physiopathology
  • Coronary Artery Disease / surgery
  • Drug-Eluting Stents / standards
  • Drug-Eluting Stents / trends*
  • Everolimus
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Single-Blind Method
  • Sirolimus / administration & dosage
  • Sirolimus / analogs & derivatives*
  • Treatment Outcome

Substances

  • Everolimus
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT00823212