The reliability of Anycheck device related to healing abutment diameter

J Adv Prosthodont. 2020 Apr;12(2):83-88. doi: 10.4047/jap.2020.12.2.83. Epub 2020 Apr 22.

Abstract

Purpose: The purpose of this in vitro study was to examine the reliability of the Anycheck device and the effect of the healing abutment diameter on the Anycheck values (implant stability test, IST).

Materials and methods: Thirty implants were placed into three artificial bone blocks with 10 Ncm, 15 Ncm, and 35 Ncm insertion torque value (ITV), respectively (n = 10). (1) The implant stability was measured with three different kinds of devices (Periotest M, Osstell ISQ Mentor, and Anycheck). (2) Five different diameters (4.0, 4.5, 4.8, 5.5, and 6.0 mm) of healing abutments of the same height were connected to the implants and the implant stability was measured four times in different directions with Anycheck. The measured mean values were statistically analyzed.

Results: The correlation coefficient between the mean implant stability quotient (ISQ) and IST value was 0.981 (P<.01) and the correlation coefficient between the meant periotest value (PTV) and IST value was -0.931 (P<.01). There were no statistically significant differences among the IST values with different healing abutment diameters.

Conclusion: There was a strong correlation between the Periotest M and Anycheck values and between the ISQ and IST. The diameter of the healing abutment had no effect on the Anycheck values.

Keywords: Implant stability; Implant stability quotient (ISQ); Implant stability test; Insertion torque value; Periotest.