Clin Chem. 1998 Apr;44(4):825-32.

Fully automated assay for total homocysteine, cysteine, cysteinylglycine, glutathione, cysteamine, and 2-mercaptopropionylglycine in plasma and urine.

Author information

  • 1Laboratory of Biochemistry, Children Hospital Bambino Gesù IRCCS, Rome, Italy.


We describe a 6-min HPLC method to measure the total concentrations of the most important thiols in plasma and urine--cysteine, homocysteine, cysteinylglycine, and glutathione--as well as the concentrations in plasma and urine, respectively, of cysteamine and 2-mercaptopropionylglycine, two compounds used to treat disorders of cysteine metabolism. Precolumn derivatization with bromobimane and reversed-phase HPLC were performed automatically by a sample processor. Throughput was up to 100 samples in 24 h. The within-run CV ranged from 0.9% to 3.4% and the between-run CV ranged from 1.5% to 6.1%. Analytical recovery was 97-107%, with little difference between plasma and urine samples. The detection limit was approximately 50 nmol/L for all the analytes studied. Thiol concentrations were determined in the plasma of 206 healthy donors and in the urine of 318 healthy donors distributed for age and sex. Mean values of plasma cysteine and homocysteine were significantly lower in infants (ages, <1 y) compared with other age groups (P <0.005). In adults, mean plasma homocysteine values were higher in males than in females (9.2 vs 6.7 micromol/L, P <0.0001) and in the 6- to 10-year-old group (P <0.05). Mean values for glutathione and cysteinylglycine were not sex- and age-dependent. In urine, both cysteine and homocysteine showed a wide range of variation.

[PubMed - indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for HighWire
    Loading ...
    Write to the Help Desk