    Eur J Clin Chem Clin Biochem. 1997 Mar;35(3):241-2.

    Evaluation of the Radiometer whole blood glucose measuring system, EML 105.

    Harff GA, Janssen WC, Rooijakkers ML.

    Department of Clinical Chemistry, Catharina Hospital, Eindhoven, The Netherlands.

    The performance of a new glucose electrode system from Radiometer was tested using two EML 105 analyzers (Radiometer Medical A/S, Copenhagen, Denmark). Results were very precise (both analyzers reported CV = 1.0% at a glucose concentration of 13.4 mmol/l). Comparison of methods was performed according to the NCCLS EP9-T guideline. Patients glucose results from both analyzers were lower compared with the results obtained with a Hitachi 911 (Boehringer Mannheim, Mannheim, Germany). There was no haematocrit dependency of relevance.

    PMID: 9127748 [PubMed - indexed for MEDLINE]