Cutting characteristics of a sonic root-end preparation instrument

Endod Dent Traumatol. 1996 Apr;12(2):96-9. doi: 10.1111/j.1600-9657.1996.tb00104.x.

Abstract

The aim of this study was to investigate the cutting ability of retro tips powered by a sonic handpiece (MM 1500, Micro Mega, Prodonta, Geneva, Switzerland). Two levels of the following variables were evaluated in this study: a) power setting with the an inlet ring half or fully open, b) orientation of tip perpendicular or parallel to the long axis of the handpiece, c) length of tip 2 or 3 min, d) loading of 25 or 50 grams, e) tip size 35 or 55. The substrate used was 1 mm thick sections of bovine bone and load was controlled by using a load cell interfaced with a transducer meter. Instrumentation time was fixed at 10 seconds with water irrigation. A 2(5) full factorial analysis was performed with two replications making a total of 64 experimental units. The resultant depth of cut was measured using a stereo microscope at x50 magnification. Analysis of the data indicated that all variables had a significant effect on cutting (ANOVA p < 0.05). The most significant factor was power, followed by tip length tip orientation, width and load. An increase in loading resulted in tip constraint and a reduction of cutting at the lower power setting. In conclusion sonically activated retro tips were found to cut satisfactorily with instrument air inlet ring opening/power having the main effect.

MeSH terms

  • Analysis of Variance
  • Animals
  • Cattle
  • Dental Instruments*
  • Equipment Design
  • Evaluation Studies as Topic
  • Femur
  • Retrograde Obturation / instrumentation*
  • Root Canal Preparation* / instrumentation*
  • Sonication / instrumentation