Ophthalmic Physiol Opt. 1996 Jan;16(1):42-8.

Contrast and luminance as parameters defining the output of the VERIS topographical ERG.

Author information

  • 1Department of Optometry and Radiography, Hong Kong Polytechnic University, Hung Hom.

Abstract

The Visual Evoked Response Imaging System (VERIS I) topographical electroretinogram system (EDI Associates, San Francisco, CA) allows measurement of the response of large numbers of retinal areas simultaneously. This paper examines ERG responses derived with the VERIS system to changes in target contrast and to local reductions in luminance. There is a linear reduction of response with reductions of target contrast. Neutral density filters were placed over part of the display, to mimic the localised reductions in response which occur in glaucoma, age-related maculopathy or diabetic retinopathy. There is a definite reduction in response seen with a 0.4 ND filter, indicating that the system should have similar sensitivity in detection of retinal lesions to the commonly used visual field analysis systems.

PMID:
8729565
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Blackwell Publishing
    Loading ...
    Write to the Help Desk