Department of Radiology, CHU Rangueil, Toulouse, France.
Under endoscopic and radiologic control, two types of self-expandable metal prostheses were implanted in tracheobronchial lesions to help reestablish airway caliber. Thirty-nine metal stent prostheses (6-20 mm in diameter) and 35 Gianturco stents (30 mm in diameter) were used in 55 adult patients with 62 lesions of the trachea (n = 33) or bronchi (n = 29). All lesions except one were endoscopically confirmed to be noninflammatory. Immediately after implantation, radiologic and endoscopic studies verified reestablishment of a satisfactory airway diameter in all patients. At a mean follow-up of 10.35 (range, 3-27) months, improvement in the respiratory status of 49 of the 55 patients (89%) was maintained and tolerance of the device was excellent. For the Wallstent endoprosthesis, the six complications observed at endoscopy were successfully treated. The Gianturco stent, however, led to a high rate of complications: 30% of cases had migration and/or rupture of the metallic mesh, potentially leading to obstruction or wall perforation; one case of respiratory distress was fatal. This procedure offers rapid epithelialization and incorporation of the device into the tracheobronchial wall.