Display Settings:

Format

Send to:

Choose Destination
See comment in PubMed Commons below
Circulation. 1993 Feb;87(2):500-11.

Augmentation of coronary blood flow by intra-aortic balloon pumping in patients after coronary angioplasty.

Author information

  • 1Department of Internal Medicine, St. Louis University, Mo.

Abstract

BACKGROUND:

Controversy exists regarding the ability of intra-aortic balloon pumping to increase coronary blood flow in patients with obstructive coronary artery disease. To assess the effects of intra-aortic balloon pumping on coronary hemodynamics, we measured coronary blood flow velocity with a 0.018-in. Doppler-tipped angioplasty guide wire in 15 patients who received an intra-aortic balloon pump for typical clinical indications.

METHODS AND RESULTS:

Intra-aortic balloon pumping augmented diastolic pressure 83 +/- 35%. In nine patients before angioplasty, peak diastolic coronary flow velocity beyond the stenosis (mean diameter narrowing, 95 +/- 7%) was 5.3 +/- 9.6 cm/sec and was unaffected by intra-aortic balloon pumping. After angioplasty, the improved coronary luminal diameter narrowing (n = 12; mean narrowing, 18 +/- 12%) was associated with increased distal diastolic flow velocity integral and peak diastolic and mean velocities (13.3 +/- 8.4 units: 36.4 +/- 18.3 and 24.0 +/- 11.4 cm/sec, respectively; all p < 0.01 versus before angioplasty), which were further augmented (36 +/- 37%, 54 +/- 49%, and 26 +/- 17%, respectively; all p < 0.01) with intra-aortic balloon pumping. Intra-aortic balloon pumping did not significantly increase the distal systolic velocity integral (10 +/- 59%) or peak systolic velocity (3 +/- 33%). Similar degrees of balloon pump augmentation of distal coronary flow velocity values were observed in five angiographically normal reference arteries in four patients.

CONCLUSIONS:

These data demonstrate lack of significant flow improvement beyond most critical stenoses with intra-aortic balloon pumping and the unequivocal restoration and intra-aortic balloon pump-mediated augmentation of both proximal and distal coronary blood flow velocities after amelioration of severe coronary obstructions in patients after successful coronary angioplasty.

PMID:
8425297
[PubMed - indexed for MEDLINE]