Send to:

Choose Destination
See comment in PubMed Commons below
J Comput Assist Tomogr. 1993 Jul-Aug;17(4):536-46.

MRI-PET registration with automated algorithm.

Author information

  • 1Department of Pharmacology, University of California, Los Angeles School of Medicine 90024-1721.



We have previously reported an automated method for within-modality (e.g., PET-to-PET) image alignment. We now describe modifications to this method that allow for cross-modality registration of MRI and PET brain images obtained from a single subject.


This method does not require fiducial markers and the user is not required to identify common structures on the two image sets. To align the images, the algorithm seeks to minimize the standard deviation of the PET pixel values that correspond to each MRI pixel value. The MR images must be edited to exclude nonbrain regions prior to using the algorithm.


The method has been validated quantitatively using data from patients with stereotaxic fiducial markers rigidly fixed in the skull. Maximal three-dimensional errors of < 3 mm and mean three-dimensional errors of < 2 mm were measured. Computation time on a SPARCstation IPX varies from 3 to 9 min to align MR image sets with [18F]fluorodeoxyglucose PET images. The MR alignment with noisy H2(15)O PET images typically requires 20-30 min.

[PubMed - indexed for MEDLINE]