[Basic study of CA125 measurement using a newly developed "SD-8729" IRMA kit]

Kaku Igaku. 1993 Oct;30(10):1245-52.
[Article in Japanese]

Abstract

"SD-8729" is a one-step IRMA kit employing OC125 antibody as 125I-labeled tracer and M-11 antibody as an immunoadsorbent. Higher bound-radioactivity to beads was observed with shorter incubation time than that of a currently used CA125 IRMA kit which employed OC125 antibody both as 125I-labeled ligand and immunoadsorbent attached to beads. Almost identical CA125 values were obtained by using two kits. The antigenic nature recognized by the M-11 antibody seems substantially different from those of 130-22 or 145-9 antibodies recognizing CA130 antigen.