Quality assurance in brachytherapy: principles for ionization chamber measurement of absorbed dose close to brachytherapy sources

Phys Med Biol. 1993 Oct;38(10):1475-83. doi: 10.1088/0031-9155/38/10/008.

Abstract

At present there exists no standardized procedure for absorbed dose determination with ionization chambers in the vicinity of brachytherapy sources. In this paper, a modification of the IAEA formalism for external photon beams is proposed in order to make it applicable at a reference point 20 mm from brachytherapy sources as well. A non-uniformity correction factor is introduced to correct for the effects of the steep absorbed dose gradient, and it is concluded that reasonable accuracy is possible when the Bragg-Gray principle is assumed to be valid.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brachytherapy / standards*
  • Humans
  • Quality Assurance, Health Care*
  • Radiometry / instrumentation
  • Radiometry / methods*