Display Settings:

Format

Send to:

Choose Destination
We are sorry, but NCBI web applications do not support your browser and may not function properly. More information
J Nucl Med. 1983 Sep;24(9):782-9.

Brain blood flow measured with intravenous H2(15)O. I. Theory and error analysis.

Abstract

The tissue autoradiographic method for the measurement of regional cerebral blood flow (rCBF) in animals was adapted for use with positron emission tomography (PET). Because of the limited spatial resolution of PET, a region of interest will contain a mix of gray and white matter, inhomogeneous in flow and in tracer partition coefficient (lambda). The resultant error in rCBF, however, is less than 4%. Although the tissue autoradiographic method requires a monotonically increasing input function to ensure a unique solution for flow, the PET adaptation does not, because of an additional integration in the operational equation. Simulation showed that the model is accurate in the presence of ischemia or hyperemia of the gray matter. Inaccuracy in timing of the arterial input function will result in large errors in rCBF measurement. Propagation of errors in measurement of tissue activity is largely independent of flow, reflecting the nearly linear flow compared with activity relationship.

PMID:
6604139
[PubMed - indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for HighWire
    Loading ...
    Write to the Help Desk