A helium gas probe for use in cryosurgery

Cryobiology. 1984 Oct;21(5):570-3. doi: 10.1016/0011-2240(84)90056-7.

Abstract

The design and testing of a prototype cryosurgical probe utilizing helium gas precooled with liquid nitrogen are described. An 8-mm-diameter probe produced an ice ball with a diameter of 28 mm after 10 min freezing using a helium gas flow rate of 42 liter/min. This indicated a surface heat transfer coefficient of 0.34 W/cm2 degrees K and temperature of -138 degrees C at the probe tip. Improved performance figures can be achieved using higher gas pressures and flow rates. A helium gas flow system schematic for use with this new type of cryoprobe is also presented. It is claimed that this system will overcome the problems of developing both multiple-tipped probes and small-diameter needle probes for use in cryoanalgesia.

MeSH terms

  • Cryosurgery / instrumentation*
  • Helium

Substances

  • Helium