Diabetes Care. 1984 Jul-Aug;7(4):378-80.

Evaluation of a self-calibrating blood glucose monitor.

Abstract

A new reflectance blood glucose meter designed for hospital use, the Reflocheck (Boehringer-Mannheim, West Germany), was evaluated. The instrument showed excellent correlation with a routine laboratory method (r = 0.996), and the coefficients of variation at high, medium, and low glucose levels were less than 4%.