Quantification of contrast in clinical MR brain imaging at high magnetic field

Invest Radiol. 1985 Jul;20(4):360-9. doi: 10.1097/00004424-198507000-00005.

Abstract

The relative contrast between two tissues in a magnetic resonance (MR) image is shown to be quantifiable for any combination of pulse timing parameters, provided the intrinsic parameters are known. Based on multiple inversion-recovery and spin echo images, a region-of-interest T1, T2 and density analysis was conducted at 1.4T in selected patients with diagnosed neuropathology for various brain tissues. The resulting tissue parameters subsequently served to calculate the contrast-to-noise (C/N) ratio for typical tissue interfaces as a function of the operator-variable pulse timing parameters and the data were compared with the images. Although such calculations may be useful as a protocol selection aid, it is obvious that an optimized pulse protocol can only be established for a single tissue interface. The data also reveal that a T2-discriminating pulse sequence like Carr-Purcell-Meiboom-Gill with long repetition time, generally advocated as clinically most effective, may not always be ideal.

MeSH terms

  • Brain Diseases / diagnosis*
  • Humans
  • Magnetic Resonance Spectroscopy* / methods
  • Multiple Sclerosis / diagnosis
  • Neuroma, Acoustic / diagnosis