Improved Applicability and Diagnostic Accuracy of the Novel Spleen-Dedicated Transient Elastography Device for High-Risk Esophageal Varices

Diagnostics (Basel). 2024 Mar 30;14(7):743. doi: 10.3390/diagnostics14070743.

Abstract

Spleen stiffness measurement (SSM) by transient elastography (TE) has been repeatedly demonstrated as the reliable way to rule out the presence of high-risk esophageal varices (HRV). We aimed to evaluate and compare novel vs. standard TE-SSM module performance in diagnosing HRV in patients with compensated advanced chronic liver disease (cACLD). This retrospective study included patients with cACLD; blood data, upper digestive endoscopy performed within 3 months of TE, SSM@50Hz and SSM@100Hz were collected. Overall, 112 patients with cACLD were analyzed (75.9% males, average age of 66, 43.7% alcohol-related chronic liver disease, 22.3% metabolic-associated steatotic liver disease, 6.2% viral hepatitis). Reliable SSM was possible in 80.3% and 93.8% of patients by using SSM@50Hz and SSM@100Hz probe, respectively. At the cut-off 41.8 kPa and 40.9 kPa (Youden), SSM@50Hz and SSM@100Hz had AUROCs of 0.746 and 0.752, respectively, for diagnosing HRV (p = 0.71). At the respective cut-offs, sensitivities for HRV were 92.9% and 100%, resulting in misclassification rates of 7.1% and 0% by using SSM@50Hz and SSM@100Hz. SSM reliably excludes HRV in cACLD patients, with measurements below 41 kPa potentially avoiding EGD in around 50% of cases, with minimal risk of HRV omission. SSM@100Hz demonstrated less measurement failures and no HRV misclassification.

Keywords: Baveno VII criteria; high-risk esophageal varices; spleen stiffness measurement; transient elastography.