Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders

J Clin Med. 2023 Oct 6;12(19):6381. doi: 10.3390/jcm12196381.

Abstract

Purpose: This study evaluated the agreement and reproducibility of ACA measurements obtained using the built-in software of the CASIA2 (Version 3G.1) and the measurements derived from expert clinicians.

Methods: Healthy volunteers underwent ophthalmological evaluation and AS-OCT examination. ACA measurements derived from automated and manual SS location were obtained using the CASIA2 automated software and clinician identification, respectively. The intraobserver, interobserver reproducibility, CASIA2-human grader reproducibility and CASIA2 repeatability were assessed using intraclass correlation coefficients (ICCs).

Results: The study examined 58 eyes of 30 participants. The CASIA2 software showed excellent repeatability for all ACA parameters (ICC > 0.84). Intraobserver, interobserver, and CASIA2-human grader reproducibility were also excellent (ICC > 0.87). Interobserver agreement was high, except for nasal TISA500, differing between observers 1 and 2 (p < 0.05). The agreement between CASIA2 measurements and human graders was high, except for nasal TISA500, where observer 1 values were smaller (p < 0.05).

Conclusion: The CASIA2 built-in software reliably measures ACA parameters in healthy individuals, demonstrating high consistency. Although a small difference was observed in nasal TISA500 measurements, interobserver and CASIA2-human grader reproducibility remained excellent. Automated SS detection has the potential to facilitate evaluation and monitoring of primary angle closure disease.

Keywords: AS-OCT; agreement; anterior chamber angle; optical coherence tomography; primary angle closure glaucoma; reproducibility.