Pulsatile flow in a catheterised artery

J Biomech. 1986;19(3):239-49. doi: 10.1016/0021-9290(86)90156-9.

Abstract

A model is presented for the physiological problem of a catheter which is inserted in a femoral artery to measure the pressure gradient. As the catheter will modify the pressure distribution in the artery, the pressure gradient which would be recorded by a perfect pressure transducer attached to it would differ from that in the uncatheterised artery. To estimate the magnitude of this error, it is assumed that the rates of flow of blood through the catheterised and the uncatheterised artery are described by the same known periodic function of time.