Display Settings:

Format

Send to:

Choose Destination
    Dtsch Med Wochenschr. 1987 Aug 28;112(35):1323-30.

    [An automatic implantable cardioverter-defibrillator. The initial clinical experience].

    [Article in German]

    Abstract

    An automatic, implantable cardioverter-defibrillator (AICD) which generates a high-energy current impulse is now available for the management of treatment-resistant malignant ventricular arrhythmias. Such a device (manufactured by Intec/CPI) was implanted into eight patients with coronary heart disease or dilated cardiomyopathy, and in four after surgery for postinfarction ventricular arrhythmias. All patients had had life-threatening episodes of ventricular fibrillation or tachycardia: the arrhythmias were refractory to multiple drug therapy (mean of 6.8 antiarrhythmia drugs per patient). The threshold energy for converting ventricular fibrillation was 9.6 Joules +/- 5.7. Except for one bacterial infection of the electrodes, there were no serious complications. During a mean observation period of 13.6 months 21 electrophysiologically induced and 105 spontaneous tachyarrhythmias were successfully terminated by the AICD. No malfunctions occurred and there was no death due to an arrhythmia.

    PMID:
    3622270
    [PubMed - indexed for MEDLINE]

      Supplemental Content

      Icon for Georg Thieme Verlag Stuttgart, New York

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk