    Clin Chem. 1987 Jun;33(6):819-20.

    Detection of interference by cyclobenzaprine in liquid-chromatographic assays of tricyclic antidepressants.

    Abstract

    We evaluated a technique for detecting cyclobenzaprine interference with liquid-chromatographic assays for tricyclic antidepressants. The technique involves dual-wavelength absorbance monitoring of the column effluent at 214 and 254 nm. Ratios of analyte peak heights at each wavelength are used to check for the presence of co-eluting interferences. With this technique, one can detect interference with an amitriptyline assay caused by 10 micrograms of cyclobenzaprine per liter.