Comparison of serum inflammatory indicators and radiographic results in MAKO robotic-assisted versus conventional total knee arthroplasty for knee osteoarthritis: a retrospective study of Chinese patients

BMC Musculoskelet Disord. 2022 May 4;23(1):418. doi: 10.1186/s12891-022-05373-y.

Abstract

Background: The purpose of this study was to compare the serum inflammatory indicators and radiographic results of conventional manual total knee arthroplasty (CM-TKA) with those of MAKO-robotic assisted total knee arthroplasty (MA-TKA).

Methods: We retrospectively analysed 65 patients with knee osteoarthritis who underwent unilateral TKA from December 2020 to November 2021 in our department, which included 34 patients who underwent MA-TKA and 31 patients who underwent CM-TKA. The tourniquet time and estimated blood loss (EBL) were compared between the two groups. Knee function was evaluated using range of motion (ROM), functional score and pain score. Leukocytes, C-reactive protein (CRP), erythrocyte sedimentation rate (ESR), interleukin-6 (IL-6), creatine kinase (CK), and neutrophil-to-lymphocyte ratio (NLR) were recorded at 3 time points (preoperative, and on the first and third postoperative days). The hip-knee-ankle angle (HKA) and the femoral and tibial component angles in the coronal and sagittal planes were used for postoperative radiographic evaluation.

Results: The postoperative MA-TKA group had less EBL (496.9 ± 257.8 vs. 773.0 ± 301.3 ml, p < 0.001). There was no significant difference in knee function scores at 6 weeks postoperatively (p > 0.05). IL-6 levels were significantly lower in the MA-TKA group on the 1st postoperative day (11.4 (5.2, 21.0) vs. 24.6 (86.3, 170.8), p = 0.031). This difference in inflammatory indices became more pronounced at 72 hours after the operation because CRP, ESR, IL-6, and CK values were significantly lower in the MA-TKA group on the 3rd postoperative day (72 h) (p < 0.05). Postoperative radiographic examinations performed 2 days after the MA-TKA group suggested that only 2 cases of HKA had outlier values, which was remarkably better than the 12 cases found in the CM-TKA group (5.9% vs. 38.7%, p < 0.001). The frontal femoral component was significantly closer to the expected value of 90° in the MA-TKA group (90.9 (90.5, 92.3) vs. 92.4 (91.3, 93.7), p = 0.031). The remaining imaging evaluation parameters were not significantly different between the two groups (p > 0.05).

Conclusions: In Chinese patients with OA, there was a milder systemic inflammatory response in the early postoperative period after MA-TKA compared to that of CM-TKA, as well as better radiographic outcomes. However, the tourniquet time was prolonged, and no advantages were observed in terms of functional score or pain score in the short-term follow-up.

Keywords: Early outcome; Inflammatory markers; Prosthesis position; Robot-assisted; Total knee arthroplasty.

MeSH terms

  • Arthroplasty, Replacement, Knee* / adverse effects
  • Arthroplasty, Replacement, Knee* / methods
  • China
  • Humans
  • Inflammation / diagnostic imaging
  • Interleukin-6
  • Knee Joint / diagnostic imaging
  • Knee Joint / surgery
  • Knee Prosthesis*
  • Osteoarthritis, Knee* / diagnostic imaging
  • Osteoarthritis, Knee* / surgery
  • Pain
  • Retrospective Studies
  • Robotic Surgical Procedures*

Substances

  • Interleukin-6