Switchable and Tunable Deep Beamformer Using Adaptive Instance Normalization for Medical Ultrasound

IEEE Trans Med Imaging. 2022 Feb;41(2):266-278. doi: 10.1109/TMI.2021.3110730. Epub 2022 Feb 2.

Abstract

Recent proposals of deep learning-based beamformers for ultrasound imaging (US) have attracted significant attention as computational efficient alternatives to adaptive and compressive beamformers. Moreover, deep beamformers are versatile in that image post-processing algorithms can be readily combined. Unfortunately, with the existing technology, a large number of beamformers need to be trained and stored for different probes, organs, depth ranges, operating frequency, and desired target 'styles', demanding significant resources such as training data, etc. To address this problem, here we propose a switchable and tunable deep beamformer that can switch between various types of outputs such as DAS, MVBF, DMAS, GCF, etc., and also adjust noise removal levels at the inference phase, by using a simple switch or tunable nozzle. This novel mechanism is implemented through Adaptive Instance Normalization (AdaIN) layers, so that distinct outputs can be generated using a single generator by merely changing the AdaIN codes. Experimental results using B-mode focused ultrasound confirm the flexibility and efficacy of the proposed method for various applications.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Data Compression*
  • Image Processing, Computer-Assisted* / methods
  • Phantoms, Imaging
  • Ultrasonography / methods