    Radiology. 1988 Sep;168(3):657-9.

    Comparison of two screen-film combinations in contact and magnification mammography: detectability of microcalcifications.

    Oestmann JW, Kopans DB, Linetsky L, Hall DA, McCarthy KA, White G, Swann C, Kelley JE, Johnson LL.

    Department of Radiology, Massachusetts General Hospital, Boston.

    Abstract

    A new dual-screen, dual-emulsion-film combination that allows a decrease in radiation dose of approximately 66% was compared with a widely used single-screen, single-emulsion-film system in contact and magnification mammography. Clustered microcalcifications randomly superimposed on a breast phantom were detected, and the location and number of individual calcifications were determined by four observers. The detectability of calcifications, determined with a receiver operating characteristic (ROC) analysis area, was 0.92 for magnification and 0.82 for contact mammography with the single-emulsion-film system, compared with 0.84 and 0.72, respectively, with the dual-emulsion-film system. More clusters were correctly located and more individual calcifications were counted with magnification than with contact mammography. The dual-emulsion-film system with the magnification technique performs as well as the single-emulsion-film system with the contact technique, while retaining a decrease in required dose of approximately 40%.

    PMID: 3406394 [PubMed - indexed for MEDLINE]