Aims: The aim of this proof-of-concept study was to investigate safety and efficacy of a CT-scan based patient-specific algorithm to maximize coronary clearance and secondarily to achieve anatomically correct commissural alignment with the Acurate Neo device.
Method and results: A total of 45 consecutive patients undergoing TAVR with the Acurate Neo THV were prospectively enrolled in the study. Mean age was 81.6 ± 5.5 years, mean STS score was 6.1 ± 3.7. Device success rate was 100%. Aim of the technique was to rotationally deploy the TAVR device with a commissure lying on the bisector between the coronary ostia as calculated on the pre-procedural CT-scan. At post-TAVR CT-scan, coronary clearance was achieved in 98% of patients with no cases of severe coronary artery overlap. In 42 out of 45 patients, THV was aligned or, at most, mildly misaligned; there were 2 cases of moderate misalignment without any case of severe misalignment. Post-TAVR selective coronary artery engagement was attempted and succeeded in all patients (100%).
Conclusion: Our CT-scan based patient-specific algorithm is safe and proven to be effective in avoiding coronary artery overlap and providing commissural alignment with Acurate Neo in all treated patients.
Keywords: TAV-in-TAV; TAVR; commissural alignment; coronary access.
© 2021 The Authors. Catheterization and Cardiovascular Interventions published by Wiley Periodicals LLC.