Revisiting prosthesis choice in mitral valve replacement in children: Durable alternatives to traditional bioprostheses

J Thorac Cardiovasc Surg. 2020 May 31:S0022-5223(20)31281-2. doi: 10.1016/j.jtcvs.2020.04.173. Online ahead of print.

Abstract

Objective: To determine risk factors for re-replacement and death or transplant following mitral valve replacement (MVR) in children METHODS: This is a retrospective 26-year review of patients younger than 20 years of age undergoing MVR between 1992 and 2018 at single institution. Outcomes included freedom from re-MVR and transplant-free survival. Cox proportional hazards regression models assessed association between outcomes and potential risk factors.

Results: At median age 4.2 years, 190 children underwent 290 MVR: 180 mechanical, 63 porcine, 13 pericardial, and 34 stented bovine jugular vein valves. Re-MVR occurred in 100 valves. Freedom from re-MVR at 5 and 10 years was 76% and 44%. Times to re-MVR were associated with prosthesis type (P < .001), with porcine and pericardial valves at greatest risk. Other risk factors for prosthetic failure included smaller prosthesis size and left ventricular hypoplasia. There were 9 transplants and 44 deaths. Transplant-free survival at 5 and 10 years was 81% and 76%. Prosthesis type was significantly associated with time to death/transplant in univariate analysis only (P = .021), with porcine at greater risk than mechanical. Independent risk factors for death/transplant included larger indexed geometric orifice area and longer bypass time.

Conclusions: In pediatric patients undergoing MVR, mechanical and stented bovine jugular vein valves were associated with increased durability compared with fixed-diameter bioprosthetic alternatives.

Keywords: Melody valve; bioprosthetic; congenital heart surgery; mechanical; mitral valve replacement; prosthetic durability.