High-Efficiency Plasma Separator Based on Immunocapture and Filtration

Micromachines (Basel). 2020 Mar 28;11(4):352. doi: 10.3390/mi11040352.

Abstract

The shortcomings of standard plasma-separation methods limit the point-of-care application of microfluidics in clinical facilities and at the patient's bedside. To overcome the limitations of this inconvenient, laborious, and costly technique, a new plasma-separation technique and device were developed. This new separation method relies on immunological capture and filtration to exclude cells from plasma, and is convenient, easy to use, and cost-effective. Most of the RBCs can be captured and immobilized by antibody which coated in separation matrix, and residue cells can be totally removed from the sample by a commercially plasma purification membranes. A 400 µL anti-coagulated whole blood sample with 65% hematocrit (Hct) can be separated by the device in 5 min with only one pipette. Up to 97% of the plasma can be recovered from the raw blood sample with a separation efficiency at 100%. The recovery rate of small molecule compounds, proteins, and nucleic acid biomarkers is evaluated; there are no obvious differences from the centrifuge method. The results demonstrate that this method is an excellent replacement for traditional plasma preparation protocols.

Keywords: microfluidics application; plasma separation; point-of-care testing; sample preparation; separator; whole blood.