Reproducibility of retinal and choroidal measurements using swept-source optical coherence tomography in patients with Parkinson's disease

Arq Bras Oftalmol. 2020 Jan-Feb;83(1):19-27. doi: 10.5935/0004-2749.20200008.

Abstract

Purpose: To assess the reproducibility of retinal and choroidal measurements in the macular and peripapillary areas using swept-source optical coherence tomography in patients with Parkinson's disease.

Methods: A total of 63 eyes of 63 patients with idiopathic Parkinson's disease were evaluated using a three-dimensional protocol of swept-source optical coherence tomography. The following layers were analyzed: full retinal thickness, retinal nerve fiber layer, ganglion cell layer, and choroid. The coefficient of variation was calculated for every measurement.

Results: In the macular area, the mean coefficients of variation of retinal thickness, ganglion cell layer + thickness, and choroidal thickness were 0.40%, 0.84%, and 2.09%, respectively. Regarding the peripapillary area, the mean coefficient of variation of the retinal nerve fiber layer thickness was 2.78. The inferior quadrant showed the highest reproducibility (coefficient of variation= 1.62%), whereas the superonasal sector showed the lowest reproducibility (coefficient of variation= 8.76%).

Conclusions: Swept-source optical coherence tomography provides highly reproducible measurements of retinal and choroidal thickness in both the macular and peripapillary areas. The reproducibility is higher in measurements of retinal thickness versus choroidal thickness.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Choroid / anatomy & histology
  • Choroid / diagnostic imaging*
  • Choroid / physiopathology
  • Choroid Diseases / diagnostic imaging*
  • Choroid Diseases / etiology*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Parkinson Disease / complications*
  • Reproducibility of Results
  • Retina / anatomy & histology
  • Retina / diagnostic imaging*
  • Retina / physiopathology
  • Tomography, Optical Coherence / methods*