[Frequency-Selectable RF Ablation Treatment System]

Zhongguo Yi Liao Qi Xie Za Zhi. 2018 Nov 30;42(6):409-412. doi: 10.3969/j.issn.1671-7104.2018.06.005.
[Article in Chinese]

Abstract

In order to support study of new radiofrequency ablation treatments and finally realize of precise conformal ablation of targeted tissue, this paper proposed a system which is able to generate radio frequency currents at different frequencies. The designed system bases on the basic principle of Class E power amplifier and uses MSP430 microcontroller as a controller and uses a touch screen as human-machine interface to design a frequency-selectable radiofrequency ablation treatment system,. The RF signals at frequencies of 230 kHz, 460 kHz, 920 kHz, and 2 000 kHz were tested to verify the feasibility of the system, and by using a tissue phantom, tested the heating ability of the RF signals at frequencies of 460 kHz, 920 kHz, and 2 000 kHz. The results show the device could well heat the tissue at the three frequencies. The study has also found that both the highest temperature and the ablation area decreases with RF frequency, when using the constant power control mode.

Keywords: Class E tuned power amplifier; RF Ablation; frequency-selectable.

MeSH terms

  • Catheter Ablation* / instrumentation
  • Hot Temperature
  • Humans
  • Hyperthermia, Induced
  • Phantoms, Imaging
  • Radio Waves*