Role of an anatomically contoured plate and metal block for balanced stability between the implant and lateral hinge in open-wedge high-tibial osteotomy

Arch Orthop Trauma Surg. 2018 Jul;138(7):911-920. doi: 10.1007/s00402-018-2918-9. Epub 2018 Mar 15.

Abstract

Introduction: Open-wedge high tibial osteotomy (OWHTO) is a well-established surgical option for medial compartment osteoarthritis of the varus knee. The initial strength of the fixation plate is critical for successful correction maintenance and healing of the osteotomy site. This study was conducted to verify if a newly designed anatomical plate (LCfit) improves the stability of both the medial implant and lateral hinge area, as well as to evaluate how the metal block contributes to both medial and lateral stability.

Materials and methods: A finite element (FE) tibial model was combined with TomoFix plate, a LCfit plate with and without a metal block. Data analysis was conducted to evaluate the balanced stability, which refers to the enforced lateral stability resulting from redistribution of overall stress. We assessed the balanced stability of the medial implant and lateral hinge area in three cases using the same Sawbones and loads using the tibia FE model.

Results: The LCfit plate reduced stress by 23.1% at the lateral hinge compared to the TomoFix plate (TomoFix vs. LCfit: 34.2 ± 23.3 MPa vs. 26.3 ± 17.5 MPa). The LCfit plate with a metal block reduced stress by 40.1% at the medial plate (210.1 ± 64.2 MPa vs. 125.8 ± 65.7 MPa) and by 31.2% (26.3 ± 17.5 MPa vs. 18.1 ± 12.1 MPa) at the lateral hinge area compared to the reduction using the LCfit plate without a metal block.

Conclusion: The newly designed fixation system for OWHTO balanced the overall stress distribution and reduced stress at the lateral hinge area compared to that using a conventional fixation system. The addition of the metal block showed additional benefits for balanced stability between the medial implant and lateral hinge area. However, this conclusion could only be drawn using the FE model in this study. Therefore, further clinical studies are necessary to reveal the clinical effect of reduced lateral stress on the occurrence of the lateral hinge fracture and the biologic effect of the metal block on the healing of the medial cortex.

Keywords: Anatomical contoured plate; Block; Finite element analysis; Knee; Open-wedge high-tibial osteotomy; Osteoarthritis.

MeSH terms

  • Aged
  • Bone Plates*
  • Equipment Design
  • Female
  • Finite Element Analysis
  • Humans
  • Knee Joint / surgery
  • Male
  • Metals
  • Middle Aged
  • Osteoarthritis, Knee / surgery
  • Osteotomy / instrumentation*
  • Osteotomy / methods*
  • Tibia / surgery*
  • Wound Healing

Substances

  • Metals